Satin-Air APP — Operation Manual
Bubble pad
• Wipe down all surfaces with soap and water before using any
other disinfectant.
• Any obvious blood spots should be wet down thoroughly with a
1:9 Hypochlorite solution (1 part bleach to 9 parts water) and allowed
to dry for at least 10 minutes. Then blot with a clean, damp cloth.
• Dry the bubble pad in a SUNLESS area after cleaning.
MAINTENANCE
General
• Regularly check the power cord and plug to inspect for abrasions
or excessive wear.
• Regularly inspect the bubble pad for signs of wear or damage.
• Ensure bubble pad and tubes are connected together correctly.
• Plug in the pump and check the airflow from the hose connection port.
Regularly inspect the air hose for any kinks or breaks.
For replacement parts, please contact your local agent or dealer.
Low pressure
If you are experiencing lower than normal pressure from your bubble pad,
check the following:
• Check connections between the air bubble pad and pump. If they are
loose or disconnected; reconnect them.
• Ensure each single cell is not broken. Set pressure dial to firm and
keep the tubes fully inflated and inspect for air leakage. If any leakage
occurs, please contact your local dealer.
HANDLING AND STORAGE
• Remove all hoses and sheets.
• Completely deflate the bubble pad.
• Lay the bubble pad out flat.
• Roll or fold the bubble pad for storage.