Guidance and manufacturer declaration – electromagnetic immunity for 
EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING

Guidance and declaration of manufacturer - Electromagnetic Immunity

For all Equipment and Systems that are not LIFE SUPPORTING

The Oxygen Concentrator is intended for use in the electromagnetic environment 
specified below. The customer or the user of the Oxygen Concentrator should ensure 
that the unit is used in such an environment.

Immunity 

test

IEC 60601 

test level

Compliance 

level

Electromagnetic  

environment - guidance

Conducted 
RF IEC 
61000-4-6

Radiated 
RF IEC 
61000-4-3

3 Vrms
150 kHz to 
80 MHz

3 Vrms
80 MHz to 
2.5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equip-
ment should be used no closer to any part of the 
OXYGEN CONCENTRATOR, including cables, 
than the recommended separation distance 
calculated from the equation applicable to the 
frequency of the transmitter.

Recommended separation distance

Where P the maximum output power rating of 
the transmitter in watts (W) according to the 
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).

Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,

a

 

should be less than the compliance level in each 
frequency range.

b

Interference may occur in the  
vicinity of equipment marked  
with the following symbol:

NOTE 1:

 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2:

 These guidelines may not apply in all situations. Electromagnetic propagation is affected 

by absorption and reflection from structures, objects and people.

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due 
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the OXYGEN CONCENTRATOR is used exceeds the appli-
cable RF compliance level above, the OXYGEN CONCENTRATOR should be observed to verify 
normal operation. If abnormal performance is observed, additional measures may be necessary, 
such as reorienting or relocating the OXYGEN CONCENTRATOR.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Содержание O2C5L

Страница 1: ...TRUAIRE 5 5 LITER OXYGEN CONCENTRATOR Manual MODEL O2C5L Read this manual before operating the concentrator Save these instructions for future reference ...

Страница 2: ...10 Unpacking Packing List Inspection Storage 8 Operating the Concentrator 10 Select a Location Setting up the Concentrator Flow Rate O2 Sensor and Alarms Turning off the Concentrator 9 Maintenance 12 Cleaning the Cabinet Cleaning the Cabinet Filter Replacing the Intake Filter Cleaning the Humidifier Bottle Cleaning and Disinfection between Patients 10 Preventive Maintenance Log 15 11 Troubleshooti...

Страница 3: ...situation which if not avoided could result in serious injury or equipment damage CAUTION Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the device or other property IMPORTANT SAFETY PRECAUTIONS AND WARNINGS It is important that you read all warnings in user s manual before use Use the product only for its...

Страница 4: ...4 LABELING SYMBOLS AND DESCRIPTIONS ...

Страница 5: ...See technical specifications for maximum storage and operating temperatures and relative humidity DISCONTINUE OPERATION IMMEDIATELY IF ANY OF THE MOTOR HOUSING PARTS BECOME DETACHED OR BROKEN EXPOSING THE MOTOR OR ANY OTHER INTERNAL ELECTRICAL COMPONENTS There is a risk of fire associated with oxygen enrichment during oxygen therapy DO NOT use the oxygen concentrator or accessories near sparks or ...

Страница 6: ... pollutants or fumes The air intake and exhaust should be well ventilated DO NOT connect the oxygen concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices ALWAYS keep oxygen concentrator at least 10 feet away from wireless communication equipment such as wi fi network devices mobile phones cordless phones and base stations walkie talkies etc DO NOT bring the o...

Страница 7: ...reation of any spark near the oxygen concentrator including static electricity Avoid use while bathing If continuous usage is required under physician order the concentrator MUST be in another room at least 10 feet from the bath or shower DO NOT come into contact with the concentrator while wet DO NOT place or store the concentrator where it can drop into water or any other liquid DO NOT reach for...

Страница 8: ...cription 1 Flowmeter 2 Status Indicator Light 3 Outlet Connector 4 Circuit Breaker 5 Power Switch 6 Hour Meter 7 Humidifier Bottle Holder 8 Power Cord 9 Cabinet Filter 10 Intake Filter 11 Caster Wheels 12 Cabinet Housing 1 9 10 11 12 7 8 3 2 4 5 6 ...

Страница 9: ... F 131 F 0 C 55 C 10 90 RH Electrical Classification Class II equipment Type BF applied part IP21 Mode of Operation Continuous Minimum Operation Time 30 minutes Safety System Low Pressure Alarm Low Oxygen Concentration Alarm High Pressure High Temperature Shutdown Current Overload Line Surge Shutdown Maximum Cannula Tubing Length 7 Cannula 50 Oxygen Supply Tubing Expected Life 6 Years Warranty Per...

Страница 10: ...e exterior of the oxygen concentrator for any visible damage including nicks dents cracks or scratches Inspect all components including accessories display panel power cord caster wheels etc STORAGE Whenever the oxygen concentrator is not in use place back in the original packaging and store in a cool dry area DO NOT place anything on top of the repackaged concentrator OPERATING THE CONCENTRATOR S...

Страница 11: ...trator to the on position After assembly ensure oxygen flow through cannula or mask NOTE Allow 30 minutes for the device to warm up to reach optimum oxygen concentration Concentrator may be used during the initial warm up time FLOWRATE Turn the flowmeter knob to the setting prescribed by your physician NOTE DO NOT set the flow above the red line on the flow meter An oxygen flow greater than 5 L mi...

Страница 12: ...ntenance the TruAire 5 Oxygen Concentrator has an expected service life of three years of operation when used in accordance with the safety instructions maintenance intervals and correct use stated in this manual The effective service life can vary according to frequency and duration of use Refer to the Preventive Maintenance Checklist in this manual for a schedule of recommended maintenance WARNI...

Страница 13: ...bricants oils or grease DO NOT service or maintain while in use DO NOT spray or apply any cleaning product directly to the cabinet DO NOT hose down the product CLEANING THE CABINET FILTER Turn off the power switch and unplug the device NOTE Conditions in the home or facility may require more frequent inspection and cleaning of the filter including high dust air pollutants pet fur and dander etc Ca...

Страница 14: ... running water and dry To disinfect the humidifier bottle and cover soak in a solution of 1 part vinegar diluted with 10 parts water Rinse thoroughly with warm running water and dry CAUTION To limit risk of bacterial growth clean the humidifier and air dry thoroughly when concentrator is not in use CLEANING AND DISINFECTION BETWEEN PATIENTS WARNING Cleaning and disinfection of the oxygen concentra...

Страница 15: ...EACH INSPECTION Date of Service Hour Meter Readout Clean Cabinet Filter Liter Flow Prescription PREVENTIVE MAINTENANCE BETWEEN PATIENTS Every 3 Years or 26 280 Hours Oxygen Concentration Replace Cabinet Filter Replace Intake Filter Check Power Loss Alarm ...

Страница 16: ...tructions blocking the fan or replace 3 Add cold water to the humidifier bottle 4 Remove excess water All repairs must be completed by an authorized repair center Concentrator is running but yel low light is on 1 Oxygen concen tration is safe but decreasing 2 Unit is over heating due to blocked air intake 1 Clean or replace filters 2 Ensure device is at least 4 inches from walls heat ers or any ot...

Страница 17: ...agnetic environment regulations RF emissions CIS PR 11 Group 1 The O2C5L Ni Oxygen Concentrator uses RF energy solely for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CIS PR 11 Class B The O2C5L Ni Oxygen Concentrator is suitable for use in all establishments including domestic and those direc...

Страница 18: ...l environment Surge IEC 61000 4 5 1 kV differen tial mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 9...

Страница 19: ...paration distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Elect...

Страница 20: ... recommended below according to the maximum output power of Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 mHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended sep...

Страница 21: ... This Warranty does not cover odors stains or fading of colors of the exterior of the unit from exposure to chemicals heat or light nor does it cover damage associated with spills This warranty shall be void and Compass Health shall be relieved of any obligation or liability associated if missing Unqualified service personnel attempt to repair the device while under warranty or there has been an a...

Страница 22: ...ass Health will assign an RMA Return Merchandise Authorization for the Dealer along with instructions for the shipping of the affected unit to a Compass Health Authorized Service Center For all returns the original purchaser must 1 Properly package the unit in the original TruAIRE 5 approved shipping carton 2 Properly display the Return Authorization Number on the shipping carton 3 Send the shipme...

Страница 23: ..._____________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ __________________________________________...

Страница 24: ...Manufactured for Compass Health Brands Corp 6753 Engle Rd Middleburg Heights OH 44130 www compasshealthbrands com Phone 800 376 7263 Made in China 42 O2C5L_00 2021 ...

Отзывы: