1 Safety Information

Do not operate this equipment until the safety information and instructions in this user guide have been read and understood by all 
personnel concerned.

The pressure envelope of the generator must not be breached under any circumstances. Failure to comply may result in an unplanned release 
of pressure, and may cause serious personal injury or death. All maintenance procedures that require the pressure envelope to breached must 
only be performed by competent personnel trained, qualified, and approved by Compair. 

Due to the nature of operation there is a possibility of oxygen enrichment surrounding the generator. Ensure that the area is adequately 
ventilated. Where the risk of oxygen enrichment is high, such as a confined space or poorly ventilated room, the use of oxygen monitoring 
equipment is advisable.

Nitrogen is not a poisonous gas but, in a concentrated form, there is a risk of asphyxiation. Depending upon the model and operating pressure, 
the generator is capable of delivering nitrogen at a flow rate of 33.3 m

3

/hr. If the generator is operated within a confined space ensure that 

adequate ventilation and oxygen monitoring equipment is fitted.

Use of the equipment in a manner not specified within this user guide may result in an unplanned release of pressure, which may cause serious 
personal injury or damage.

When handling, installing or operating this equipment, personnel must employ safe engineering practices and observe all related regulations, 
health & safety procedures, and legal requirements for safety.

Ensure that the equipment is depressurised and electrically isolated, prior to carrying out any of the scheduled maintenance instructions 
specified within this user guide.

Only competent personnel trained, qualified, and approved by Compair should perform installation, commissioning, service and repair 
procedures.

Note: Any interference with the calibration warning labels will invalidate the gas generator’s warranty and may incur costs for the re-calibration 
of the gas generator.

Compair can not anticipate every possible circumstance which may represent a potential hazard. The warnings in this manual cover the most 
known potential hazards, but by definition can not be all-inclusive. If the user employs an operating procedure, item of equipment or a method of 
working which is not specifically recommended by Compair the user must ensure that the equipment will not be damaged or become hazardous 
to persons or property.

Most accidents that occur during the operation and maintenance of machinery are the result of failure to observe basic safety rules and 
procedures. Accidents can be avoided by recognising that any machinery is potentially hazardous.

Details of your nearest Compair sales office can be found at www.compair.com

Retain this user guide for future reference.

Содержание CN20033

Страница 1: ...ZS 176510020 000 07 2012 EN Original Language Nitrogen Gas Generator CN20033 CN20090 User Manual...

Страница 2: ...ce the company shall not be under any liability whatsoever under the warranty if before giving notification in writing to the company as aforesaid the Customer or any third party meddles interferes ta...

Страница 3: ...ion 10 3 3 1 Generator Supply 11 3 3 2 Dryer Supply 11 3 3 3 Purge Economy 11 3 3 4 Remote Switching 11 3 3 5 Alarm Contacts 11 3 3 6 4 20mA Analogue Output 11 4 Operating the Generator 12 4 1 Overvie...

Страница 4: ......

Страница 5: ...se serious personal injury or damage When handling installing or operating this equipment personnel must employ safe engineering practices and observe all related regulations health safety procedures...

Страница 6: ...ntrol and Laboratory use Part 1 General Requirements EN 61000 6 1 2007 Electromagnetic compatibility EMC Part 6 1 Generic standards Immunity for residential commercial and light industrial environment...

Страница 7: ...1 0 5 1 2 3 4 5 Flowrate CN20033 m3 hr 0 55 1 2 1 5 1 9 2 4 3 4 4 3 5 8 7 2 8 4 9 4 cfm 0 3 0 7 0 9 1 1 1 4 2 0 2 5 3 5 4 2 4 9 5 5 CN20072 m3 hr 1 2 2 4 3 2 3 9 4 7 6 9 8 5 11 6 14 3 16 7 18 8 cfm 0...

Страница 8: ...98 11 92 408 16 32 682 27 28 754 30 16 869 34 76 203 8 12 233 9 32 98 216 1 CN20072 1034 41 36 450 18 640 25 6 375 15 467 18 68 408 16 32 682 27 28 754 30 16 869 34 76 203 8 12 233 9 32 145 319 7 CN20...

Страница 9: ...in a clean dry environment If the crate is stored in an area where the environmental conditions fall outside of those specified in the technical specification it should be moved to its final location...

Страница 10: ...Mains supply cable gland 2 Column A pressure gauge 8 O2 Analyser calibration port 3 Column B pressure gauge 9 N2 Outlet port to buffer G1 2 4 N2 Outlet pressure gauge 10 N2 inlet port from buffer vess...

Страница 11: ...0 particles in the 0 1 0 5 micron size range are allowed In each cubic metre of compressed air not more than 6 000 particles in the 0 5 1 micron size range are allowed In each cubic metre of compresse...

Страница 12: ...yer pre filtration 7 Generator 10 Dust filter 2 Wet air receiver 5 Pre treatment dryer 8 Buffer vessel 11 Ball valve 3 Water separator 6 Dust filter 9 Pressure relief valve 12 Drain valve Flowrate Ves...

Страница 13: ...the outlet port of the buffer vessel The AR010 dust filter provided should be installed in this line Follow the installation instructions provided with the filter taking note of the direction of flow...

Страница 14: ...L Grey Dryer Live conductor 3 7 mm N Blue Dryer Neutral conductor Yellow Green Dryer Earth conductor JP 17 Purge Economy JP17 1 NC Not Used JP17 2 COM The relay is energised when the generator is in s...

Страница 15: ...remote alarm circuit The contacts are rated 1A max 250Vac 1A 30Vdc Under normal operation the relay is energised when a fault occurs the relay will de energise causing the relay contacts to change st...

Страница 16: ...ted to generators without an integrated O2 analyser provides a visual indication of the operating status The liquid crystal display gives a reading of the total number of hours that the generator has...

Страница 17: ...BV3 and check for leaks in the piping between the vessel and the generator 7 Open the ball valve BV4 on the N2 Outlet port Note If the purity of the gas is not within specification applies only to gen...

Страница 18: ...an be used to maintain the beds when the generator is in economy mode If the over ride is enabled a clean up cycle will be performed once every 20 minutes This allows the generator to go straight on l...

Страница 19: ...ly default back to the main operating menu if no key activity has been detected for one minute After an additional two minutes of inactivity the display will shutdown To bring the display back on line...

Страница 20: ...password 1 2 1 _ _ When the password has been entered correctly the Hour Meters menu will be displayed Use the key to navigate to page 3 Customer Settings The time in hours that the generator has bee...

Страница 21: ...ion Sets the purity level at which an oxygen fault is initiated Default Settings If the purity level exceeds the oxygen alarm level for a period longer than the alarm delay the oxygen alarm will be ac...

Страница 22: ...4000 Hours 6 month 8000 Hours 12 month 16000 Hours 24 month Generator Check status indicators located on the control panel Generator Check regulated outlet pressure Generator Check O2 purity System Ch...

Страница 23: ...a Compair trained qualified and approved engineer Description Kit No Kit MIST X Silencer 606280162 Description Kit No Kit Filter Element 1x ZS1087938 Description Kit No Kit PPM Oxygen Cell 1x 6064000...

Страница 24: ...the sensor as detailed Note PPM cells require the offset value to be entered prior to calibrating 5 4 3 Dust Filter Element Change C Close the ball valves located on the inlet and outlet ports of the...

Страница 25: ...21 C E F D 8 9 10 10 11 A B...

Страница 26: ...o stabilise before entering the calibrated level Using compressed air Select menu 3 1 and enable the O2 Alarm Over ride Connect the O2 sample line between the elbow push in fitting located on the ball...

Страница 27: ...lve back to its original position and remove the regulated calibration gas supply or independent analyser as applicable If the calibration is not successful the original reading from the analyser will...

Страница 28: ...Name Address Telephone Fax Contact Name Date of Commission Service Interval Months Hours Date Serviced By Comments Print Sign 6 4 000 12 8 000 18 12 000 24 16 000 30 20 000 36 24 000 42 28 000 48 32...

Страница 29: ...h Oxygen concentration Defective Oxygen cell Replace the oxygen cell Leak in system piping Open the access door and check all connection points for leaks Repair as necessary Low inlet pressure The pre...

Страница 30: ...TITLE REVISION 5 R S 10 P N 15 20 25 M L K J 30 35 40 45 50 55 R S P N M L K J H G F E D C B A 5 10 15 20 25 H G F E 30 35 40 D C 45 50 B A 55 006510005 Basic Schematic E...