Measuring AC and DC current

7

WARNING: To avoid damage to the meter or 
injury if the fuse blows, never attempt an 

in-circuit current measurement where the 

open-circuit potential to earth (ground) is 

greater than 250V.

The DC current range is 400uA, 4.000mA, 
40.00mA, and 400.0mA .The AC current 
range is 400uA, 4.000mA, 40.00mA, 
and 400.0mA.

NOTE: Connect the measuring current of 
the meter in series, not connected in 
parallel. This prevents damage to the 
meter or endangering personal safety.

 Cut off the power of the measured circuit. 

    

 Discharge all high voltage capacitance on 

    

 the measured circuit.

 Turn the rotary switch to the mA or uA 

    

 position (1). When the measured current 

    

 is less than 400uA, select the uA position. 

   

 

 

When the measured current is 

    

 4mA~400mA, select the mA position.

 Disconnect the circuit under test. Connect 

    

 the black test lead to the end of dis

    

 connected circuit under test (with lower 

    

 voltage). Connect the red test lead to the 

    

 other end of the disconnected circuit 

    

 under test (with higher voltage).

 Switch on the power of the circuit, and then 

    

 read the reading displayed on the LCD. If the 

    

 display only displays “OL”, it indicates that 

    

 the input exceeds the selected range. Move 

    

 the rotary dial to a higher range.

 Cut off the power to the measured circuit. 

    

 Discharge all capacitances, take off the 

    

 test leads and recover the circuit.

When measuring DC current, if the test 
leads are reversely connected to the 
circuit , the display will change into 
negative, but not affect measuring 
accuracy.

17

16

Operation (continued)

Measuring capacitance

Operation (continued)

5

CAUTION: To avoid the meter or measured 
equipment from being damaged, cut off all 

power supply of measured circuits and 

discharge all high voltage capacitors before 

on/off measurement.

The capacitance ranges of the meter are 

4.000nF, 40.00nF, 400.0nF, 4.000uF, 

40.00uF, 100.0uF.

 Turn the rotary switch to the  

    

position (1).

 Press the FUNC key (2) and select the 

    

CAP measuring range.

 Apply the two ends of the test leads to 

    

measure the two pins of the capacitance 

    

under test and read the measured value 

    

on the LCD.

NOTE: The meter will take some time to 
stabilize the reading when the capacitance 

is high. A small capacitance with less than 
10nF subtracts from the distribution 
capacitance of the meter and lead 

 

(namely 

displayed base number) when measuring.

Measuring frequency and duty cycle

6

 Turn the rotary switch to the  

    

 position (1).

 Press the Hz% key (2). For frequency, 

    

 select the Hz range. For duty cycle, 

    

 select the % range.

 Apply the two ends of the test leads to 

    

 measure the frequency or duty cycle 

    

 value of the circuit under test.

 Read the value that displays on the LCD.

Содержание 730 696

Страница 1: ...MS8332C 730 696...

Страница 2: ...his multimeter and the full scale AC 220V protection design allows for safe operation This series of meters can be widely used for schools labs research institutes enterprises and factories Please rea...

Страница 3: ......

Страница 4: ...0313 or visit www homedepot com WARRANTY 12 months PACKAGE CONTENTS Pre Operation Part Name Quantity A B Multimeter Black test lead C Red test lead Non contact AC Voltage Indicator Light LCD Display A...

Страница 5: ...mpedance 10M Maximum Input Voltage 600V DC AC VOLTAGE SPECIFICATIONS Used to select functions and desired ranges Select Switch Switching key between frequency and duty cycle it is used for switching H...

Страница 6: ...pen Circuit Voltage Diode is approximately 1 5V buzzer is approximately 0 5V Overload Protection 250V DC or rms AC CAPACITANCE SPECIFICATIONS Pre Operation continued Measuring Range Resolution Accurac...

Страница 7: ......

Страница 8: ...y OL For on line measuring diode the meter displays the forward voltage drop at forward measurement the reverse measurement depends on the values of other elements connected in parallel at both ends o...

Страница 9: ...range Cut off the power to the measured circuit Discharge all capacitances take off the test leads and recover the circuit When measuring DC current if the test leads are reversely connected to the ci...

Страница 10: ......

Отзывы: