IE-SNC-T4211-01 

 

  5 

Technical parameters: 

 

Analog output: 

Range from 0 to 10 V 

   

Output load capability: min. 20 kΩ 

   

Voltage output in case of error: approximately –0.1 V or >10.5 V 

Measuring parameters: 

Temperature probe: Pt1000/3850 ppm connected by shielded cable of maximum 

length 10m 

Measuring temperature range: -200 to +600 °C (can be limited by applied 

temperature probe model) 

Resolution: 0.1 °C 
Accuracy without probe:  output ± (0,15 + 0,1% FS) °C, 

 ±0,2 °C for value on LCD display  

Recommended interval of calibration

: 2 years 

Measuring interval and LCD display refresh

: 0.5 s 

Communication with computer

: via USB port by means of USB communication  

cable SP003 

Power:

 15 to 30 V dc, maximum ripple 0.5 %, maximum load 20mA 

Protection

:  IP65 

Operating conditions: 

Operating temperature range: -30 to +80 °C, over +70°C switch LCD display OFF 
Operating humidity range: 0 to 100 %RH 
Outer characteristics in accordance with Czech National Standard 33-2000-3:  

normal environment with the specifications: AE1, AN1,  BE1 

Working position: arbitrary 
Electromagnetic compatibility: complies EN 61326-1  
Not  allowed  manipulations:  It  is  not  allowed  to  operate  the  device  under  other  than 

specified  conditions  in  technical  parameters.  Devices  are  not  designed  for 
locations with chemically aggressive environment.  

Storing conditions

: temperature -30 to +80 °C 

humidity 0 to 100 %RH without condensation 

Dimensions

: see dimensional drawings

 

 

Weight:

 approximately 145 g 

Material of the case

: ABS 

 
 
 

 

Error States of the device 

Device  continuously  checks  its  state  during  operation.  In  case  error  is  found  LCD  displays 

corresponding error code: 

 

Error 0 

First line displays „Err0“. Check sum error of stored setting inside device’s memory.  Output 

value is <-0.1 V. This error appears if incorrect writing procedure to device’s memory occurred or 
if damage of calibration data appeared. At this state device does not measure. It is the serious error, 
contact distributor of the instrument to fix. 