After connecting the logger to the computer reading of logger info is enabled by means of the 

PC software and also setting of the instrument accordingly to the user needs (menu 

Configuration / 

Setting of instrument  parameters

). Before logging start it is necessary: 

 

 

check or optionally set the logger real time clock 

 

select suitable logging interval 

 

select logging mode (cyclic or non cyclic) 

 

set for each channel type of temperature probe (Pt1000 or Ni1000) 

 

disable channels for logging with not connected temperature probe 

 

switch ON the logger (or switch OFF, if it is about to be switched ON by magnet or 
automatically with delayed start) 

 

enable or disable the option to switch ON the logger by magnet 

 

enable or disable the option to switch OFF the logger by magnet 

 

enable or disable the option to clear minimum and maximum value memory by magnet 

 

set date and time of logger automatic switching ON logger or disable this option 

 

select if record runs permanently or only if alarm is active 

 

If alarms are about to be applied, set both limits for each measured quantity and enable 
alarm 

 

optionally enable permanent alarm indication (alarm with memory) 

 

switch ON or switch OFF the display of logger 

 

optionally switch ON displaying of MIN/MAX values on the LCD 

 

reset memory of MIN/MAX values (if needed) 

 

check free space in data memory, optionally erase data memory of the logger 

 

enter password if protection against unauthorized manipulation with the logger is necessary  

 
Logging interval between to subsequent measurements is specified by the user. Memorizing of the 
first value is synchronized with the internal real time clock, so the logging is performed at sharp 
multiples of minutes, hours and days. E.g. after starting the logging with the 15 minute interval the 
first value is not stored immediately, but after the internal clock gets the status of a quarter, a half or 
a whole hour. After starting the logging with the 6 hour interval the first value is stored at that 
whole hour to perform the storing also at 00.00, i.e. at the beginning of the day. First storing is 
performed at 6.00,12.00, 18.00 or 00.00hour - at the hour from the above nearest to the logging 
start. After communication with computer or after start by magnet logger automatically waits for 
the nearest whole multiple of time and then first measurement is performed. This is also necessary 
to take into account when setting time of automatic logger switch ON. 

Notice:

 If logger operates as permanently connected to the computer, using of magnet 

start/stop is disabled. 

 To enable the logger control by magnet is suitable only in cases, when possibility of 

unauthorized manipulation to the logger operation is eliminated. 

 

Reading on the display in usual operation (logger switched ON) 

 
After switching ON the logger all LCD symbols are displayed for checking 
the display. 
 
 
Then actual date and time in logger is displayed for approximately 4 s. 
 
 

4  

IE-LOG-S0141-04 

Содержание LOGGER S0141

Страница 1: ...LOGGER S0141 Four channel temperature logger with display Instruction Manual...

Страница 2: ...written by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm limits Stored values can be transferred from logger memory to the PC by m...

Страница 3: ...tion by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation it is not allowed to operate instrument with not connected temperature probes if all probes are not connected unc...

Страница 4: ...f the logger enter password if protection against unauthorized manipulation with the logger is necessary Logging interval between to subsequent measurements is specified by the user Memorizing of the...

Страница 5: ...al temperature Notice if alarm is active no MIN MAX reading is displayed If display is switched OFF all above readings are displayed up to estimated remaining battery capacity and then display goes ou...

Страница 6: ...New initialization of logger can occur in switching ON the logger immediately after displaying of all LCD segments for checking e g after replacement of totally discharged battery for new one State i...

Страница 7: ...magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line After dec...

Страница 8: ...after its life is necessary to liquidate ecologically Connection of female connector ELKA 3008V to temperature probe It is possible to order as an optional accessory connector for connection of temper...

Отзывы: