Comet Commeter D3121 Скачать руководство пользователя страница 12

 
 

 

 

 

12

 

 

IE-COM-D3121-13

 

Logging mode 

 

 

In  logging  mode  measured  values  are  stored  to  instrument  internal  memory.  Memory  is  non 
volatile, i.e. if battery voltage drops or battery is disconnected, stored data will not be lost. 
 
Device can work in two logging modes: 

 

Automatic mode

 

– actual measured values are stored to memory in adjustable time interval. 

Storing of the first value is synchronized with internal real time clock, so the memorizing is 
performed  at  sharp  multiples  of  minutes,  hours  and  days.  E.g.  if  the  storing  interval  is 
15 minutes  value  will  be  stored  each  15.,  30.,  45.  a  00. minute  and  zero  second  in each 
hour,  if  interval  is  3  hours  values  will  be  stored  in  00:00:00,  03:00:00,  06:00:00.  etc.  It 
means  after  starting  automatic  logging  instrument  waits  for  the  nearest  integer  multiple  of 
time and then stores the first value.  

Manual  mode

 

–  actual  measured  value  is  stored  after  pressing  ENTER  key  (instrument 

should be in default mode of displaying actual measured values). Each storing is indicated 
by  a  short  beep  and  position  of  the  record  in  memory  is  displayed  (reading  MEMO  with 
the record serial number). In this mode it is possible to store max. 1000 records.  

 

It is possible to select one of the three logging modes (MODE): 
 

Automatic  non  cyclic  logging

,  when  logging  stops  after  fulfilling  the  memory.  Reading 

MEMO FULL is displayed (reading disappears after pressing any key) 

 

Automatic cyclic logging

, when after fulfilling the memory oldest data are replaced with new 

values  

 

Manual logging

, designed for manual storing of measured values by ENTER key  

  
Logging  modes  are  not  mutually  compatible  and  after  changing  logging  mode  memory  of  the 
instrument should be cleared. 

 

 

Logging  in  progress  will  always  be  interrupted  by  instrument  connection  to  the  computer  or 
switching OFF the instrument by ON/OFF key. Logging in progress will be automatically stopped if 
battery  voltage  below  7  V  is  indicated  by  the  instrument.  This  prevents  to  log  values  effected  by 
low battery voltage.  
 

 

 

 

Logging start from device keyboard 

 
 

Start logging by confirmation of the menu item START (symbol Strt) 
by pressing ENTER key. 
 
Notice:  it  is  not  possible  to  start  logging  if  low  battery  voltage  is 
indicated. 

 

 

 

 

 

Содержание Commeter D3121

Страница 1: ...www cometsystem com Instruction manual Commeter D3121 Digital thermometer hygrometer and datalogger...

Страница 2: ...values and minimum a maximum memory 9 Functions and settings available from menu 9 Alarm indication and setting 11 LOGGING MODE 12 Logging start from device keyboard 12 Logging start with clearing me...

Страница 3: ...Then it is possible to transfer data to a personal computer Temperature is measured by RTD sensor Ni1000 6180ppm Instrument compares measured values of temperature humidity and dew point with two adju...

Страница 4: ...approximately 1m s temperature step 20 C Relative humidity Measuring range 0 to 100 RH Resolution 0 1 RH Accuracy 2 5 RH from 5 to 95 RH at 23 C Response time t90 1 min with plastic sensor cover witho...

Страница 5: ...umulator In usual operation from adapter accumulator is charged only with small current If accumulator is totally discharged its full charging in instrument takes approximately 100 hours Instrument wi...

Страница 6: ...______________________________________________________________________________________________________________________________________________________________ Backlighting turns off automatically 10...

Страница 7: ...Not allowed manipulation It is not allowed to touch the sensors under the cover to prevent damage the sensors The sensors under the cover should not be exposed to direct contact with water or other l...

Страница 8: ...n is finished instrument starts to display internal time to warn to incorrect internal time New time adjustment is performed by the computer Press MENU key to switch instrument to actual measurement m...

Страница 9: ...isplay minimum resp maximum value of other values Pressing MIN MAX or MENU key again to return to default mode Minimum and maximum memory is cleared from menu after confirmation selection CLR see belo...

Страница 10: ...s displayed This value illustrates battery condition Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode DYN...

Страница 11: ...of the arrow keys Press and hold the arrow key UP to make value increase fast Press and hold the arrow key DOWN to make value decrease fast Release the arrow key and press ENTER to confirm new limit R...

Страница 12: ...dicated by a short beep and position of the record in memory is displayed reading MEMO with the record serial number In this mode it is possible to store max 1000 records It is possible to select one...

Страница 13: ...indicated with reading LOG on display If memory is filled over 90 symbol LOG is blinking only in automatic logging Logging start with clearing memory and logging mode selection It is possible to clear...

Страница 14: ...de is selected after confirmation by ENTER key logging is active In automatic logging mode i e cyclic or non cyclic last applied storing interval is offered to confirm or modify See above procedure Lo...

Страница 15: ...protocol of measurement including instrument serial number and a table of logged values with date and time After disconnection of the communication cable reading PC disappears and instrument is in the...

Страница 16: ...t for automatic Installator run If Installator is not run automatically depends on setting of operating system find in file browser file CDSetup exe in main directory of the installation CD and click...

Страница 17: ...le of measured values occurs Detailed information on file is available from context menu menu occurs after clicking right mouse button on the position of table In context menu function export to text...

Страница 18: ...cessary to replace battery with instrument switched OFF otherwise internal time and setting of d REF and AUDI from menu selections and data in memory HOLD MIN and MAX will be lost For the same reason...

Страница 19: ...IE COM D3121 13 19 Technical support and service Technical support and service is provided by distributor Contact is included in warranty certificate...

Отзывы: