, s.r.o., Bezrucova  2901  

 

Specifications are subject to change without notice. 

756 61 Roznov pod Radhostem, Czech republic 

IE-SNC-T0210-09

 

Instruction manual for use of T0210 transmitter 

Transmitter  is  designed  for  measurement  of  ambient  air  temperature  at  °C  or  °F  and  relative 

humidity  of  air  without  aggressive  ingredients  with  computing  one  of  the  following  values:  dew 
point  temperature,  absolute  humidity,  specific  humidity,  mixing  ratio  and  specific  enthalpy. 
Measuring  temperature  and  relative  humidity  sensors  are  non-removable  instrument  parts. 
Measured  and  computed  values  are  displayed  on  dual  line  LCD  display.  The  first  line  displays 
temperature.  Value  displayed  on  the  second line  is  selectable  among  relative  humidity  and 
computed  value.  It  is  also  possible  to  display  both  readings  with cyclic  overwriting  in  4  seconds 
interval. It is possible to switch OFF the LCD at all. It is possible to assign measured or computed 
value  to output  Uout1  or  output  Uout2.  Both  voltage  outputs  have  common  ground  with  power 
source (GND terminal). 

All  transmitter  setting  is  performed  by  means  of  the  PC  connected  via  the  optional  SP003 

communication  cable  (not  included  in  delivery).  Program 

Tsensor

  for  transmitter  setting  is 

available  to  download  free  at 

www.cometsystem.com

.  Program  enables  to  assign  each  output 

measured value (temperature, relative humidity, computed value) and its range. It supports make the 
adjustment of the device too. This procedure is described at file „

Calibration manual.pdf

“ which is 

installed commonly  with the software.  It  is  also  possible to  assign both  outputs  to  the same value 
(with the same range), if two evaluation devices are necessary to connect.

  

Transmitter  version  T

xxxx

L

  with  watertight  male  connector  instead  of  a  cable  gland  is 

designed  for  easy  connection/disconnection  of  the  output  cable.  The  protection  of  male  Lumberg 
connector RSFM4 is IP67. 

Models  marked  T

xxxx

Z

  are  non-standard  versions  of  the  transmitters.  Description  is  not 

included in this manual. 

Please read instruction manual before the first device connection. 

 

Device setting from the manufacturer 

Transmitter is set from the manufacturer to the following parameters: 

value at output Uout1:

 relative humidity, range 0 - 10 V corresponds 0 to 100 % RH 

value at output Uout2:

 temperature, range 0 - 10 V corresponds -30 to +80 °C 

display:

 switched ON 

value displayed at line 2:

 relative humidity only 

 

Modification of the setting is possible to do by means of the PC using procedure described 

at the end of this document. 

Installation of the transmitter 

Transmitter  is  designed  for  wall  mounting.  There  are  two  mounting  holes  at  the  sides  of  the 

case.  It  is  NOT  recommended  to  use  the  device  for  long  time  under  condensation  conditions.  It 
could  be  the  cause  of  water  steam  condensation  inside  the  sensor’s  cover  into  water  phase.  This 
liquid phase stays inside sensor’s cover and can’t escape from the cover easily. It can dramatically 
increase response time to relative humidity change. If water condensation occurs for longer time it 
can  cause  sensor  damage.  Similar  effect  can  occur  under  water  aerosol  conditions.  Don’t  connect 
transmitter  while  power  supply  voltage  is  on.  Interconnection  terminals  of  T0210  are  accessible 
after unscrewing four screws and removing the lid. Lace the cable through a gland at the case wall. 
Connect  the cable to  terminals  with  respecting the signal polarity (see  figure). Terminals  are self-
clamping  and  can  be  opened  by  a  suitable  screwdriver.  For  the  opening,  insert  the  screwdriver  to 
upper terminal hole and lever by him. Do not remember to tighten glands and case lid with inserted 
packing  after  cables  connecting.  It  is  necessary  for  warranting  of  protection  IP65.  Connect 

Содержание T0210

Страница 1: ...yright COMET SYSTEM s r o It is prohibited to copy and make any changes in this manual without explicit agreement of company COMET SYSTEM Ltd All rights reserved COMET SYSTEM Ltd makes constant develo...

Страница 2: ...of a cable gland is designed for easy connection disconnection of the output cable The protection of male Lumberg connector RSFM4 is IP67 Models marked TxxxxZ are non standard versions of the transmi...

Страница 3: ...ble must be located at indoor rooms The cable should not be led in parallel along power cabling Safety distance is up to 0 5 m otherwise undesirable induction of interference signals can appear Outsid...

Страница 4: ...Type of value assigned to output 1 is indicated by displayed unit here RH relative humidity Upper line displays voltage value corresponding to measured value lower line Press button again to get value...

Страница 5: ...wn are for value displayed on LCD display For value on analog output is valid too if selected output range is set inside measuring range Analog outputs Two outputs with range from 0 to 10 V with commo...

Страница 6: ...s Operating temperature range 30 to 80 C over 70 C switch LCD display OFF Operating humidity range 0 to 100 RH Outer characteristics in accordance with Czech National Standard 33 2000 3 normal environ...

Страница 7: ...ine displays Err0 Check sum error of stored setting inside device s memory Output value is 0 1 V This error appears if incorrect writing procedure to device s memory occurred or if damage of calibrati...

Страница 8: ...r appeared converter does not respond probably damage of A D converter No measurement and calculations of values are proceeded Output value is about 0 1 V It is a serious error contact distributor of...