Specific humidity
1
Accuracy:
±2g/kg at ambient temperature T < 35°C (95 °F)
Range:
0 to 550 g/kg
Mixing ratio
Accuracy:
±2g/kg at ambient temperature T < 35°C (95 °F)
Range:
0 to 995 g/kg
Specific enthalpy
Accuracy:
± 3kJ/kg at ambient temperature T < 25°C (77 °F)
Range:
0 to 995 kJ/kg
2
The values computed from ambient temperature and relative humidity including their
accuracy you can exactly determine by the program
Conversions
. It is free to download at
9.2.
Operating conditions
Operating temperature range:
electronics
H5021
:
-30 to +80 °C (-22 to 176 °F)
electronics
H5024, H6020
:
-30 to +60 °C (-22 to 140 °F)
measuring end of stem
H6020:
-30 to +80 °C (-22 to +176 °F)
CO
2
probe
H5021:
-40 to +60 °C (-40 to +140 °F)
It is recommended to switch off the LCD display at ambient temperatures above 70°C
(158 °F)
around electronics.
Operating humidity range:
H5021:
0 to 100 % RH (no condensation)
H5024, H6020:
5 to 95 % RH (no condensation)
Operating pressure range:
850 to 1100 hPa
Protection:
H5021
IP65 (electronics), IP65 (CO
2
probe)
H5024
IP65 (electronics)
H6020
IP30 (electronics), IP40 (measuring end of stem)
Other environmental conditions (external influences) according to HD 60364-5-51:
normal
Recommended calibration interval:
H5021
5 years (CO
2
)
H5024
5 years (CO
2
)
H6020
5 years (CO
2
), 1 year (relative humidity), 2 years (temperature)
Working position:
H5021
negligible
H5024
with cable glands upwards
H6020
with sensor cover downwards
1
This value depends on the atmospheric pressure. For computing is used constant value stored in device memory. Default
value preset by manufacturer is 1013hPa and can be changed by user’s software.
2
This maximum is reached under conditions about 70°C/100%RH or 80°C/70%RH