Specifications 

8

 

 

Humidity Input 

- moisture-content range 
- relative-humidity range 
- measurement error 

Temperature Input 

- temperature range 
- measurement error 

Supply Voltage 
Batteries 
 
Operational Consumption 
Stand-by Consumption 
Non-linearity 
Temperature Drift 
Voltage-to-Humidity Influence 
Voltage-to-Temperature Influence 
Ambient Temperature / Humidity 
Protection Class 

from RHP probe 
0...300 % 
0...75 % 
0.2% from span 

 Pt1000 (IEC751) 

-20…150 °C 
0.1% from span 
3 V from 2 batteries AA 1.5 V 

 no batteries, 

 regular batteries,  

 long-life, 

 rechargeable  

less than 8 mA 
less than 1 

µ

negligible 
0.01% from span for 1 °C 
0.2% from span for 1 V  
0.1% from span for 1 V  
-10…65 °C / 0…85% RH 
IP44 

 

 

Warranty and Support 

 

 

…………………………. 

           serial number 

 

…………………………. 

      manufacturing date

 

 

QC check mark …….……(passed)   

(stamp) 

88 Slavyanska Str. 
P.O.Box 378 
Plovdiv 4000, BULGARIA  
tel: +359 32 646523, 646524 
fax: +359 32 634089, 646517 
e-mail: [email protected]      

 

QD-8.2.4-WC

 

 

 

 

 

 

 

 

Warranty

 

COMECO warrants this product  
to be free from defects in materials  
and workmanship for 2 years. If your unit 
is found to be defective within that time,  
we will promptly repair or replace it.  
This warranty does not cover accidental 
damage, wear or tear, or consequential  
or incidental loss. This warranty 
does not cover any defects caused by 
wrong transportation, storage, installation,  
or operating (see '

Specifications

').  

Technical support

 

In the unlikely event that you encounter  
a problem with your COMECO device, 
please call your local dealer or contact 
directly our support team. 

 

 

 

COMECO Inc., P.O.Box 378, Plovdiv 4000, BULGARIA, tel: +359 32 646523, 646524, fax: 634089  

e-mail: [email protected], WWW.COMECOGROUP.COM 

 

 

 

 

 

H5850 

 

O PER A T ION  MAN U AL 

 

 

 

 

 

Please read this Operation Manual before mounting and operating! 

Save the Manual for future references! 

 

 

 

RPH specifications 

6

 

 

Sensors 
  - humidity 
  - temperature 
Measurement Ranges  
  - moisture-content / RH 
  - temperature 
Output Signal (humidity) 
Power Supply 
Measurement Error 
Temperature Drift 

Cable Length / Insulation 
Operating (Ambient) Temperature 
Ambient Humidity  
Protection Class 

 
capacitive 
Pt1000, class B (IEC751) 
 
0...300 % / 0...75% 
-10...50 °C 
min. 0...max. 1 V DC  
from H5850 handheld meter 

 ± 2.5 % from span 

0.1 % from span for 1 °C 

.............. m  /  

 PVC, 

 SL 

-10…50 °C 
0...75% RH 
IP65 

 

 

Operation with RPH probe 

 

 

 

 

Before applying new material always 
clean and dry the internal probe 
surface. 

 

Fill the probe with the material to its 
upper edge and flat the surface.  

 

Do not overfill or under fill! 

Do not press or stuff the material!

 

 

For calibration purposes use two 
samples with enough different 
moisture content, but with the same 
size of the particles. 

 

In order to achieve the required 
measuring accuracy, the composition 
and the particles' size of the material 
being tested must be the same as the 
calibration samples. 

 

 

Operation 

3

 

 

 

 

Notes

 

1. If the resolution is integer (no 

decimal point), H5850 will 
round off the readings with a 
decimal part greater than or 
equal to 0.5   (e.g., if the 
measured value is 12.7, 
 the unit will indicate

 13

)! 

2

. After recovery from power 

supply failure (for example, 
due to battery replacement), 
H5850 will display the 
readings without decimal 
point (integer) until the other 
resolution is selected! 

 

 

Notes

 

1. Using probes without 

Pt1000 sensor does not 
allow temperature 
measurement with H5850! 

2. The default measuring 

channel (after power-on) is 
always moisture/RH! 

 

Turning ON/OFF 

 

Turn on by pressing the 

 

key once.  

The test string 

8888

 appears on display 

and H5850 enters measurement mode.  

 

Turn off by pressing the 

 

key once. 

The unit indicates 

0FF

 and turns itself off. 

Display resolution selection 

Display resolution can be changed alternatively 

with 

 The reading can be displayed either 

as integer number or as number with one digit 
after the decimal point. 
After pressing the key, read for half a second 
respectively

 0

 or 

)0

Changing measured channel 

Besides the material humidity, H5850 can also 
measure  its  temperature,  if  the  probe  is 
equipped  with  Pt1000  temperature  sensor. 

Press once 

key to check which channel is 

being  measured  -  moisture/RH  (

AH.o

 

/

rH.o

)  or  temperature  (

t  .C

/

t  .F

). 

Press and hold 

 to change the channel. 

Measurement mode 

In measurement mode H5850 measures 
material moisture/RH or temperature depending 
device settings and connected probe and.   
One measurement cycle lasts approximately 
0.5 s and display readings refresh after its end. 

Low-battery indication 

H5850 monitors the battery voltage, and when it 
drops bellow the critical (approx. 2.5V), each 10 
seconds the device displays

 

bAtt

 for about 

0.5 s. It reminds you to replace the batteries. 

 