![COM-power corporation CLCE-452 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/com-power-corporation/clce-452/clce-452_instruction-manual_2638784015.webp)
Page 15 of 18
INSTRUCTION MANUAL
SECTI ON 4
-
CALIB RATION
Rev090519
The insertion loss of a current probe, at any given frequency, quantifies the
difference between the voltage quantity present on the center conductor
of the test fixture into which the probe is installed and the voltage quantity
present at the output connector of the probe.
To determine the insertion loss of the probe, a known voltage is applied into
one side of the calibration fixture, with the opposite side terminated into 50
ohms; while measuring the voltage present at the output of the probe. The
difference between the applied voltage and the measured voltage, at
any given frequency, is the insertion loss of the probe.
Shown in Figure 9 is the equivalent schematic circuit for the calibration. The
physical test setup showing the equipment and connections is illustrated in
Figure 10.
As measured in either of the circuits shown in Figure 9, the following
equation defines the insertion loss of the probe, at any given frequency:
NOTE:
Typical Insertion Loss Values for the CLCE-452 are represented in Figure 5
(Section 3.7).
Содержание CLCE-452
Страница 2: ...Page 2 of 18 INSTRUCTION MANUAL TABLES OF CONTENTS Rev090519...
Страница 3: ...Page 3 of 18 INSTRUCTION MANUAL TABLES OF CONTENTS Rev090519...
Страница 5: ...Page 5 of 18 INSTRUCTION MANUAL SECTION 2 PRODUCTS AVAILABLE FROM COM POWER Rev090519...
Страница 11: ...Page 11 of 18 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev090519...
Страница 12: ...Page 12 of 18 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev090519...
Страница 13: ...Page 13 of 18 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev090519...