Item

Description

A

IEC Power Entry Module / Line Cord

B

T6.3A (5 × 20 mm) Fuse – Do Not Substitute

1.  Place the power switch in the off position.

2.  Disconnect the AC power input line cord from the receptacle.

3.  Remove and check the fuse and replace if defective.

Figure 4-1. 

Fuse Replacement

A

B

B

A

Содержание Masterflex 07591-20

Страница 1: ...ECISION VARIABLE SPEED MODULAR DRIVES Model Nos 07591 20 07591 30 A 1299 1084 Edition 01 1 800 MASTERFLEX 627 8373 U S and Canada only 11 847 549 7600 Outside U S 847 549 7600 Local www masterflex com...

Страница 2: ...on Variable Speed Modular Drives Operating Manual ii Preface 2013 Thermo Fisher Scientific Inc All rights reserved NORPRENE PHARMED and TYGON Reg TM Saint Gobain Performance Plastics Corp Trademarks b...

Страница 3: ...TION To avoid electrical shock the power cord protective grounding conductor must be connected to ground Not for operation in wet locations as defined by EN61010 1 If the product is not used in a mann...

Страница 4: ...uit that will shut the motor down if any of the following conditions exist a Tubing that is too hard is loaded in the pump b Incorrect tubing size or wall thickness is loaded in the pump c Tubing is i...

Страница 5: ...ection and Replacement 3 1 Pump Controls 3 1 Keypad Lockout Enable Disable 3 2 External Operation 3 2 External Inputs 3 3 Transmitter Type 3 4 MAINTENANCE 4 1 Replacement Parts 4 1 Fuse Replacement 4...

Страница 6: ......

Страница 7: ...d Connectors 1 2 DB 9 and NEMA 4X Pin Configuration with Wiring Scheme 3 2 Typical 4 20 mA Wiring 3 4 Fuse Replacement 4 2 Motor Brush Replacement 4 3 I P Precision Variable Speed Modular Drives Opera...

Страница 8: ......

Страница 9: ...id applications Low maintenance sealless and valveless design Valveless design prevents clogging Inner surfaces are smooth and easy to clean Contamination free fluid contacts only the tubing or tube m...

Страница 10: ...rnal Local operation from the front panel keypad is designated by INT external Remote operation is designated by EXT In INT mode START STOP FLOW DIRECTION and SPEED keys on the front panel determine o...

Страница 11: ...diameter and as short as possible WARNING Turn drive off before removing or installing tubing Fingers or loose clothing could get caught in drive mechanism Use a tube size of appropriate diameter for...

Страница 12: ......

Страница 13: ...e pump to the outside of the occlusion bed to the suction side of the pump This will avoid excessive tubing wear at any specific point Always move the worn tubing to the suction side of the pump CAUTI...

Страница 14: ...white 3 Speed Control Input Reference Common 5 orange 4 Local Remote Speed Control 14 red yellow 5 Local Remote Speed Control Reference 10 grey 6 Start Stop and CW CCW Reference 8 yellow 7 Start Stop...

Страница 15: ...anel is desired with external Start Stop contact operation the INT EXT key must again be in the EXT position In addition Jumper A should be in place Jumper A connects pin 4 pin 14 for NEMA 4X models o...

Страница 16: ...ng Manual Cole Parmer Section 3 Operation Figure 3 2 Typical 4 20 mA wiring Transmitter Type ISAType 2 Two wire Transmitter Receiver Power Supply ISAType 3 Three wire Transmitter Receiver Power Supply...

Страница 17: ...ption Part Number Brushes set of 2 07520 04 Brush Cap Holder 07520 03 Fuse T6 3A 5 x 20 mm 77500 24 Line Cord Australia 50001 60 Line Cord Denmark 50001 62 Line Cord India 50001 64 Line Cord 115V Unit...

Страница 18: ...se Replacement Item Description A IEC Power Entry Module Line Cord B T6 3A 5 20 mm Fuse Do Not Substitute 1 Place the power switch in the off position 2 Disconnect the AC power input line cord from th...

Страница 19: ...hould be checked every 6 months or 2 000 operating hours 1 Disconnect AC line cord from the AC receptacle 2 Disconnect motor cable from the controller 3 Remove the two vinyl caps covering the brush ca...

Страница 20: ......

Страница 21: ...ot blown located in the slot next to the power cord Verify the motor is properly connected to the controller Check the tubing Tubing should be snug but not tight against the rollers Verify that the mo...

Страница 22: ......

Страница 23: ...20 only 1 Footswitch w DB 9 male 77595 35 2 DB 9 Remote Control Connector 07595 45 3 DB 9 Remote Control cable 25 FT 07595 47 4 Motor Extension cable 9 FT 07557 51 Model 07591 30 only 1 Washdown Foots...

Страница 24: ......

Страница 25: ...0 10V DC 10 kohm Accuracy 0 5 Full Scale Current input 4 20 mA 250 ohm Accuracy 0 5 Full Scale Environment Operating Temperature 32 to 104 F 0 to 40 C Storage Temperature 13 to 149 F 25 to 65 C Humidi...

Страница 26: ...S plastic and vinyl 07591 30 Powder coated aluminum and steel Enclosure Rating 07591 20 IP33 per IEC 60529 07591 30 IP66 NEMA 4X per IEC 60529 Compliance UL 61010 1 CAN CSA C22 2 No 61010 1 This produ...

Страница 27: ...distributor s most recent catalog This warranty shall not apply to a defects or damage resulting from i misuse of the product ii use of the product in other than its normal and customary manner iii ac...

Страница 28: ...ns before returning the product either within or outside of the warranty period When returning the product please state the reason for the return For your protection pack the product carefully and ins...

Страница 29: ......

Страница 30: ...Printed in U S A 1 800 MASTERFLEX 627 8373 U S and Canada only 11 847 549 7600 Outside U S 847 549 7600 Local www masterflex com techinfo coleparmer com...

Отзывы: