background image

20

9.3.   Servicing.

This product should be serviced by a Cole-Parmer Service Engineer or by an agent
on behalf of Cole-Parmer Limited. If in doubt contact Cole-Parmer. See Section 11.

9.4.   Calibration

9.4.1.  Recalibration

• Have a precision reference thermometer of calibrator to hand. Ensure that OS1=0 
and SIM=0
• Switch the controller off and on again.
• During the auto-test phase, press buttons                +                and keep them 
pressed until the controlled shows 0AD.

• With buttons               and                selected 0AD or SAD;0AD allows a calibration 

of 0, inserting a constant corrective over the while scale of measurement. SAD allows 
a calibration of the top part of the measurement scale with a proportional correction 
between the calibration point and 0
• Press               to display the value and then use               +                or               to 
make the read value coincide with the value measured by the reference instrument.
• Exit from calibration by pressing button               .

9.5.   Spillage and Decontamination.

In the event of spillage switch off and unplug this product from the mains electrical
supply. Wipe off all excess liquid from the unit and surrounding area using an
absorbent soft cloth.

If in doubt please consult Customer Support. Refer to section 11.

 

If the equipment has been exposed to contamination, the    
Responsible Body is responsible for carrying out appropriatede  
contamination. If hazardous material has been spilt on or inside the 

equipment, decontamination should only be undertaken under the control of 
the Responsible Body with due recognition of possible hazards. Before using 
any cleaning or decontamination method, the Responsible Body should check 
with the manufacturer the proposed method will not damage the equipment.

Prior to further use, the Responsible Body shall check the electrical safety of 
the unit. Only if all safety requirements are met can the unit be used again. 
The above procedure is intended as a guide. Should spillage occur with a toxic 

or hazardous fluid then special precautions may be necessary.

Decontamination Certificate.

 Note:

 In the event of this equipment or any part of the unit becoming damaged, or 

requiring service, the item(s) should be returned to the manufacturer for repair

accompanied by a completed decontamination certificate. 

Copies of the Certificate 

are available from Distributor/Manufacturer.
At the end of life, this product must be accompanied by a completed 

Decontamination Certificate.

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

 

 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off. 

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The 

 

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to 
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

 

 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to 
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The 

 

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

 

 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The 

 

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming 

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Page 20 of 24

 M6493 Issue 10.3 

9.3. Servicing.

This product should be serviced by a Bibby Scientific Service Engineer or by an agent 
on behalf of Bibby Scientific l. If in doubt contact Bibby Scientific. See Section 10.

9.4. Calibration

9.4.1.  Recalibration

Have a precision reference thermometer of calibrator to hand. Ensure that

OS1=0 and SIM=0

Switch the controller off and on again.

During the auto-test phase, press buttons button’s

 and 

keep them pressed until the controlled shows 0AD.

With buttons

 and

 selected 0AD or SAD;0AD allows a 

calibration of 0, inserting a constant corrective over the while scale of
measurement. SAD allows a calibration of the top part of the measurement 
scale with a proportional correction between the calibration point and 0

Press 

 o display the value and then use

+

 or

 to make the read value coincide with the value measured by the 

reference instrument. 

Exit from calibration by pressing button

9.5. Spillage and Decontamination.  

In the event of spillage switch off and unplug this product from the mains electrical
supply. Wipe off all excess liquid from the unit and surrounding area using an 
absorbent soft cloth. 

If in doubt please consult Customer Support. Refer to section 10.

  If the equipment has been exposed to contamination, the 

Responsible Body is responsible for carrying out appropriate 
decontamination. If hazardous material has been spilt on or inside the 
equipment, decontamination should only be undertaken under the control of
the Responsible Body with due recognition of possible hazards. Before using
any cleaning or decontamination method, the Responsible Body should check 
with the manufacturer the proposed method will not damage the equipment.

Prior to further use, the Responsible Body shall check the electrical safety of
the unit. Only if all safety requirements are met can the unit be used again. The 
above procedure is intended as a guide. Should spillage occur with a toxic or
hazardous fluid then special precautions may be necessary. 

Decontamination Certificate.

Note: In the event of this equipment or any part of the unit becoming damaged, or
requiring service, the item(s) should be returned to the manufacturer for repair
accompanied by a decontamination certificate. 

Copies of the Certificate are

available from Distributor/Manufacturer. Appendix A of this instructions book
may be copied and enlarged.  

At the end of life, this product must be accompanied by a Decontamination
Certificate. See section 6.3 and 6.4

If in doubt please consult Customer support. Refer to section 10.

Page 12 of 24

 M6493 Issue 10.3 

7.3.3.  Proportional, Integral and Derivative Control.

A proportional controller continuously adjusts the power, so the heat input to 
the process is approximately in balance with the process heat requirements
in order to maintain a stable temperature. The range of temperature over
which power is adjusted is from 0% to 100%. This temperature range is
called the ‘Proportional Band’. The proportional band is adjustable for
differing process conditions (zero proportional band returns the controller to 
basic on / off control).

The temperature difference between the stabilised temperature and the set 
point is called the ‘Offset’. The offset can be reduced by adjusting the integral
term of the PID controller. This is achieved using the integrative action time
and integrative action reset. 

A derivative function can be combined with the proportional controller to 
provide the controller with the ability to shift the proportional band either up to 
or down to compensate for rapidly changing temperature (i.e. when an oven 
door is opened or cooler fluid is introduced into a controlled vessel). The 
derivative function increases the controller gain during temperature changes, 
and can help to reduce overshoot on start-up.

Hence, a three-mode PID controller combines the proportional, integral and 
derivative actions, to allow the set point temperature to be approached
relatively smoothly with minimal overshoot.

7.4. Front Panel Controls

The MC810B controller has four front panel keys as described below.

7.4.1.  The 

key is used to increase the value of the parameter shown on 

the display or to scroll through the parameter menu. 

7.4.2.  The 

 

key is used to decrease the value of the parameter shown on

the display or to scroll through the parameter menu. 

7.4.3.  The

key is a dual function key. It  is used to display and store the 

value of the selected parameter when in ‘Programming’ mode. When used with

the

 it is used to display the ‘Setpoint’ value.

7.4.4.  The 

is an ‘Exit / Stand-by button. Press of 3 seconds to turn unit on

or off.

7.5. Operating Modes. 

7.5.1.  The MB810B has two modes of operation: Function mode and Programming

mode. The unit enters Function mode immediately following power up in auto
test phase. In Function mode the display shows the temperature measured by
the temperature sensing probe. The output power will also be switched on or off
as appropriate. If the output power is on, the amber neon indicator on the front 
panel will be illuminated. The only parameter which can be altered in the 
function mode is the set-point. See 7.7.1

7.5.2.  In Programming mode, all the other parameters apart from set-point can be 

altered. To gain access to the programming mode from Function mode, press

the keys

and

. Keep them pressed for five seconds. The 

output power is always off when in programming mode. If approximately 25 
seconds elapse while in programming mode with out a key being pressed the 

Содержание Electrothermal MC810B

Страница 1: ...Digital Heating Controller MC810B MC810BX1 Instruction Manual M6493 Version 10 5...

Страница 2: ...ection 13 EC Declaration of Conformity 23 The copyright of this instruction manual is the property of Cole Parmer Limited This instruction manual is supplied by Cole Parmer Limited on the express unde...

Страница 3: ...atinum resistance probe The sample temperature is displayed on the LED display This product is suitable for bench and retort stand mounting or wall mounting using the bracket and screws provided 1 3 T...

Страница 4: ...ify Setpoint 2 button Exit Standby button Alarm This feature is not available on MC810B 4 Page 4 of 24 M6493 Issue 10 3 2 SYMBOLS AND USING THIS INSTRUCTION BOOK 2 1 Throughout this Instruction book t...

Страница 5: ...uct so that it is difficult to disconnect from the mains supply Do not immerse unit in water or fluids Do not spill substances onto this unit If spillage does occur disconnect unit from mains supply a...

Страница 6: ...supply It is not recommended to leave any heating apparatus unattended during operation Only use Original Equipment manufacture s spares and accessories Ref Section 11 Page 5 of 24 M6493 Issue 10 3 En...

Страница 7: ...plug and lead set May be different from illustration As Req 2 MC810B MKII Digital Controller 1 3 Instruction Book 1 4 Wall fixing Bkt x 1 and screws x 2 per bag 1 5 Temperature Probe 1 For future refe...

Страница 8: ...ot 5 8 If the controller is to be used on a retort rod make sure it isn t positioned over any heating apparatus 5 9 If using the wall mounting bracket provided carefully mark and drill two suitable ho...

Страница 9: ...issues within the design and manufacturing process without compromising end product performance and value 6 2 Packaging materials have been selected such that they may be sorted for recycling 6 3 At...

Страница 10: ...te label and ensure your mains electrical supply voltage is compatible with this product Page 10 of 24 M6493 Issue 10 3 7 PRODUCT OPERATION 7 1 The MC810B controller has been designed for easy operati...

Страница 11: ...vely large and the heater will cause overshoot and wide variations in temperature 11 Page 11 of 24 M6493 Issue 10 3 7 2 Connection 7 2 1 Plug in the temperature probe to the 5 pin DIN socket Place the...

Страница 12: ...er This is achieved using the integrative action time and integrative action reset A derivative function can be combined with the proportional controller to provide the controller with the ability to...

Страница 13: ...is a dual function value of the selected parameter when in the it is used to display the S 7 4 4 The is an Exit Stand by b or off 7 5 Operating Modes 7 5 1 The MB810B has two modes of oper mode The un...

Страница 14: ...used to increase the the display or to scroll through the paramete 7 4 2 The key is used to decrease th the display or to scroll through the paramete 7 4 3 The key is a dual function key value of the...

Страница 15: ...r menu 7 4 2 The key is used to decrease the value of the param the display or to scroll through the parameter menu 7 4 3 The key is a dual function key It is used to display value of the selected par...

Страница 16: ...ad variations 1PF ON OFF Output state in case of probe failure OAU NON THR AL0 AL1 AUX output operation NON output disabled always off the next perameter will be ATM THR output programmed for second t...

Страница 17: ...bling INP 0mA 4mA Sensor input selection see table of input specifications RLO 19 9 RHI Maximum range values RLO takes the maximum value measured by the transmitter i e the value matching 1V 20mA RHI...

Страница 18: ...earthed ground 2 meters long Moulded IEC plug and Lead set supply cord SJT VW 1 Replace only with equivalent cable Lead set plug fuse UK only 13A BSEN1362 Mains Output Non detachable 3 core mains cab...

Страница 19: ...remove it from the mains supply Open fuse draw and remove fuses Fit replacement fuses and close the draw 9 MAINTENANCE 9 1 General Information Unplug the unit from the mains voltage supply before und...

Страница 20: ...o increase the value of the parameter shown on the display or to scroll through the parameter menu 7 4 2 The key is used to decrease the value of the parameter shown on the display or to scroll throug...

Страница 21: ...your local sales specialist or email cpspares coleparmer com to enquire about available spares Please visit www electrothermal com for a full list of available accessories CRM6322 11 CUSTOMER AND TEC...

Страница 22: ...22 12 NOTES...

Страница 23: ...hat other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that injury damage or loss might...

Страница 24: ...Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Email cpinfo coleparmer com Web www electrothermal com...

Отзывы: