displayed value should change.
2.05 Press the
CAL
key. Note that the annunciator LED next to the key is
flashing. Press the
Gear Set
key to exit out of CAL.
2.06 Verifying the LED for each tubing illuminates, press the
tubing SIZE
switch to get the
RPM
display.
2.07 Verifying each LED illuminates for
mL
,
COPY
, and
SEC
, press the
DISPense/copy
key until none of the LED's is on.
2.08 Turn the drive
OFF
. While holding the
MODE
switch, turn the drive
ON
.
Set the input to
4-20mA
(115 Vac) or
0-20mA
(230 Vac) with the
increment
(
^
) or decrement (
v
) keys. Press the
MODE
switch. Set the
output
to
4-20mA
(115 Vac) or
0-20mA
(230 Vac), with the increment (
^
)
or decrement (
v
) keys. Press the
MODE
switch to get to the
INT
mode.
Turn the drive
OFF
then
ON
after a wait of at least three seconds. Verify
the
INT
and
RPM
indicators remain lit.
2.09
Press
START
and then press
^
until display reads the maximum setting.
Using the hand tach, verify motor runs within the specified tolerance.
NOTE: Motor should be rotating CW when viewed from the front.
2.10
Press
v
until display reads the minimum setting. Using the hand tach,
verify motor runs within the specified tolerance.
Maximum Speed
Minimum Speed
Drive
Setting Max. Min. Setting
Max.
Min.
75211-30, -35,
-40, -45
3600 3618 3582 60
69
51
2.11 Press and hold the
PRIME
key while noting that the drive runs at full
speed.
2.12
Press
the
Remote Start/Stop
(NEMA drives only) or foot switch. Verify
that the drive runs as long as it is held. Press
STOP/START.
2.13
Press
the
Remote Prime
switch (NEMA drives only). Verify that the drive
runs at full speed as long as it is held.
2.14 Turn the controller
OFF
and unplug it from the variac.
**************************
TESTS COMPLETED
***********************