3-2
Economical Gravity Convection Oven
Cole-Parmer
Section 3
Unpacking and Installation
Installation
1. Install the shelves on the rack inside the oven.
2. Turn the power switch OFF and plug the oven into a receptacle that
provides the required voltage as indicated on the unit’s nameplate.
3. Install an immersion thermometer (furnished by the user) through the
thermometer grommet and into the thermometer holder.
4. Adjust the immersion depth to 6-7/8” (175mm), measured from the
bulb to the top of the rubber washer.
5. Insert this assembly into the thermometer vent ring that is located on
top of the oven.
6-7/8"
175 mm
(Top of Oven)
Thermometer Vent Ring
Thermometer Holder