6 .5 Common Commands
*CAL?
Calibrate Self-Test and Return Status
*CLS
Clears the instrument settings
*IDN?
Identification String (model and serial number)
*RST
Reset instrument
*TST?
Self-Test Trombone and Return Status
*OPC
Set Operation Complete Bit
*OPC?
Query the Operation Complete Bit
*CAL?
Calibration Self-Test Check? (QUERY)
Description
Performs a calibration self-test check on the electromechanical trombone and returns status
Returns
0
Calibration self-test passed.
1
Calibration self-test failed. Service Required.
*CLS
Clear Status (QUERY)
Description
Clears the instrument and resets any pending errors codes.
Returns
None
*IDN?
Instrument Identification Information (QUERY)
Description
Returns the Instrument identification information.
Returns
Colby Instruments,XT-100-625P,11111234,V1 .00
*RST
Reset instrument (QUERY)
Description
Performs an instrument device reset to its initial power-on state. Delay is set to zero delay setting in all
channels.
Network IP addresses (if previously changed) will take effect
Returns
None
*TST?
Calibration Self-Test Check (QUERY)
Description
Performs an internal self-test by moving the electromechanical trombones to the maximum range limit,
testing the optical sensor detection, returning to zero delay position, and re-initializing all internal variables.
Returns
0
Internal self-test passed.
1
Internal self-test failed.
26