Operation Note
12
4.
Set
the
HP
8502A
Input
Attenuation
knob
to
0
dB.
5.
Adjust
the
HP
8754A
trace
to
the
top
graticule
line
using
the
CH
1
REFERENCE
POSITION
knob.
Turn
the
CH
1
REFERENCE
POSITION
off.
6.
On
the
HP
8754A,
turn
on
the
CH
1
REFERENCE
OFFSET.
Adjust
the
trace,
reference
line,
to
the
top
CRT
graticule
using
the
REFERENCE
OFFSET
knob.
The
HP
8754A
is
now
calibrated
for
making
return
loss
measurements
relative
to
the
short.
7.
If
an
HP
8750A
Storage
Normalizer
is
used,
press
the
following
keys
on
the
HP
8750A:
[CH
1],
[INPUT],
[STORE
INPUT],
and
[INPUT
‐
MEM].
The
displayed
reference
line
will
shift
to
the
middle
of
the
screen.
Errors
in
the
reflection
test
set
will
now
be
normalized.
8.
Remove
the
SMA
Coaxial
Short
and
replace
it
with
the
Cokeva
11940C
probe.
The
displayed
trace
should
be
16
dB
or
more
below
the
reference
line
at
all
frequencies.
If
the
trace
is
less
than
16
dB
below
the
reference
line,
the
probe
fails
the
test.
5
‐
1.
SMA
ROTARY
JOINT
INSTALLATION
PROCEDURE
Figure
10.
SMA
ADAPTER
LOCATION
Figure
11.
RF
Rotary
Joint
Mounted
on
Probe