Cojali JT2005 Скачать руководство пользователя страница 2

JT2005

Infrared Thermometer
Operation Manual

Data Storage 

In data storage mode (with new data in HOLD state), press SET, data flashes 3 times and is saved in current data 

group, then the screen will display the data of next group. If no data has been saved in the group, ‘- - - -’ appears. You 

can short press MODE to return to view the previous stored data.

Delete Data 

In data storage mode, long press SET for 5s to delete all data (99 groups).

Scheduled Measurement 

Enable this function to measure at a preset time (in 24 hours format). The thermometer will turn on automatically to 

perform the measurement and save the record at preset time.

General measurement mode: Long press MODE for 3s to enter the scheduled measurement ON/OFF setup interface (as 

shown below), use LOG or MODE button to switch between ON/OFF. Thermometer will exit the setup after 5 seconds of 

no operation.

Scheduled Measurement Time Setup 

In the scheduled measurement ON/OFF setup interface, short press SET to enter the scheduled measurement time setup 

interface (as shown below), press LOG/MODE to set the hour, then short press SET and use LOG/MODE to set the minutes.

Monitoring Measurement 

Enable this function to setup the monitoring measurement at a preset interval (1 min-24h). The Thermometer will 

automatically turn on to measure and record a measurement at every interval.

In the scheduled measurement time setup interface, short press SET to enter the monitoring measurement setup 

interface (as shown below), press LOG or MODE button to switch between ON/OFF. Thermometer will exit the setup 

after 5 seconds without any operation.

Monitoring Measurement Time Setup 

In the monitoring measurement ON/OFF setup interface, short press SET to enter the monitoring measurement time 

setup interface (as shown below), press LOG/MODE to set the hour, then short press SET and press LOG/MODE to set 

the minutes.

Protection Function 

Without protection function, the thermometer will always be on during locked measurement mode, even when 

trigger is enabled by accident. When protection function is enabled, the thermometer will stop the measurement and 

shut down when there is no button operation for 10 minutes.

In the monitoring measurement lime setup interface, short press SET to enter the protection mode ON/OFF setup 

interface (as shown below), use LOG or MODE button to switch between ON/OFF. Thermometer will exit the setup 

after 5 seconds without any operation.

System Time Setup 

Please set the system time before the scheduled measurement and monitoring measurement. This thermometer 

clock is displayed in 24-hour time format.

In the protection mode ON/OFF setup interface, short press SET to enter the system time setup interface (as shown 

below), press LOG/MODE to increase/decrease the relevant value, and short press SET to set the year-month-date-

hour-minute accordingly.

Thermometer will exit the setup after 5 seconds of no operation.

Notes:

Enter the scheduled measurement/monitoring measurement/protection function/system time setting modes

The records will be saved from the 000 position when they are stored in memory. After all 99 positions are 

occupied, the recording will return to 000 again and overwrite the original data.

When this function is turned on and the thermometer is in the SCAN or HOLD states, “Auto” or “Interval” will be 

displayed.

When the thermometer restarts, the system time and the measurement settings will be restored to the default 

values (it needs to be reset), but the recording memory will be still saved.

It is not recommended to enable the scheduled measurement and monitoring measurement at the same time, as 

their measurement records all start from 000, and will overwrite the measurement data of each other.

In the HOLD state, press 

MODE

 button for more than 3s to enter the scheduled measurement mode; then press SET 

to enter the following

setting interface:

1.

2.

3.

4.

In HOLD state, press MODE (>3s)

Scheduled measurement on/off (“Auto” and “LOG” blue characters appear)

Set the hour of scheduled measurement (hour characters flash)

Set the minutes of scheduled measurement (minute characters flash)

Monitoring measurement on/off (“Interval” and “LOG” blue characters appear)

Set the hour of the interval in monitoring measurement (hour characters flash)

Set the minutes of the interval in monitoring measurement (minute characters flash)

Protection function on/off (“Protect” blue characters appear)

Set the year of the system date (year characters flash)

Set the month of the system date (month characters flash)

Set the date of the system date (date characters flash)

Set the hour of the system time (hour characters flash)

Set the minutes of the system time (minute characters flash)

Save and exit the setup mode

Press SET

Finding Hot Point and Cold Point 

Aim at target area with thermometer, then 

move up and down slowly to sweep the whole 

area until the hot point and cold point are 

found.

D:S Ratio (Distance to spot ratio) 

S increases as D increases (S represents 90% of energy in the 

circle). The maximum D:S can be obtained when D=800mm 

(32”) and S=40mm (1.6”).

Field of View 

Ensure that the target is larger than S. The smaller the target 

is, the closer the measurement distance should be. Optimum 

result is obtained when diameter of the target is more than 

2.0 times of S. The spot diameter is approximately equal to the 

distance between the two laser points.

Emissivity 

Emissivity characterization reflects the radiated energy by the material. Emissivity for most organic materials, paints or 

oxidized surfaces are about 0.95. Total emissivity of selected metals and non-metals are listed in the following table.

Maintenance

Clean

Use clean compressed air to blow away falling particles.

Use a wet cotton swab to carefully wipe lens surface.

Use a wet sponge or a soft cloth to clean product exterior.

Battery Replacement

Install or replace 1604A 9V battery according to the picture on the right.

Metal

Ceramic

0.95

Clay

0.95

Concrete

0.95

Cloth

0.9

Plates

0.96

Plaster

0.8 - 0.95

Ice

0.98

Limestone

0.98

Paper

0.95

Plastics

0.95

Water

0.93

Soil

0.9 - 0.98

Wood

0.9 - 0.95

Aluminum

      Oxidization

0.2 - 0.4

Hastelloy

      Alloy

0.3 - 0.8

Iron (Forging)

      Passivation

0.9

Molybdenum

      Oxidization

0.2 - 0.6

Nickel

      Oxidization

0.2 - 0.5

Platinum

      Black

0.9

Zinc

      Oxidization

0.1

A3003 Alloy

      Oxidization

      Rough

0.3

0.1 - 0.3

Brass

      Burnishing

      Oxidization

0.3

0.5

Inconel

      Oxidization

      Sand-Blasting

      Electro Burnishing

0.7 - 0.95

0.3 - 0.6

0.15

Iron (Casting)

      Oxidization

      Non-Oxidization

      Casting

0.6 - 0.95

0.2

0.2 - 0.3

Steel

      Cold Rolling 

      Steel Plate Rubbing

      Steel Plate Burnishing

0.7 - 0.9 

0.4 - 0.6

0.1

Non-Metal

      Asbestos

      Asphalt

      Basalt

0.95

0.95

0.7

Glass

      Convex Glass

      Smooth Glass

      Lead-Boron Glass

0.76 - 0.8

0.92 - 0.94

0.78 - 0.82

Copper

      Oxidization

      Electric Terminal Board

0.4 - 0.8

0.6

Iron

      Oxidization

      Rusting

0.5 - 0.9

0.5 - 0.7

Lead

      Rough

      Oxidization

0.4

0.2 - 0.6

Measured Surfaces

Emissivity

Long press 

MODE for 3s.

Long press 

SET for 5s.