Getting Started: Stand-off Installation
Attach the stand-off to your DataMan 7500 Series reader.
Getting Started: Calibration
Start the DataMan Setup Tool and connect to your reader.
Select the
Code Quality Settings
task pane.
Place the verifier against the calibration card with the symbol
centered below the reader, and squeeze the trigger.
1.
2.
3.
Continue squeezing the trigger. Calibration can take several seconds.
4.
The Setup Tool displays a graphic guide if the symbol is not precisely
aligned with the reader; adjust the position and rotation of the reader
until the center of the red rectangle lies within the green rectangle and
there is no more than 10 degrees rotation between the symbol and the
reader. The reader beeps once when calibration is complete.