background image

should also be replaced if any part of it is cracked. Air can leak in through the crack and cause 
errors in the readings. 

SAMPLE PROBE ASSEMBLY 

The probe assembly is a one piece assembly and cannot be taken apart. The tubing used is a 
very narrow bore inert polymer which extends from the flanged end to the tip of the probe 
assembly.  The inert composition prevents O2 adsorption on the internal surface. The internal 
tubing volume is small to minimize the volume of sample required to flush out the sensor to 
provide an adequate response time. The tip is a male luer fitting which will accept any luer-
hubbed needle or filter. When screwed into the bulkhead fitting, the captive black screw 
presses the flanged end tightly against a mating piece, also flanged, which connects to the 
inlet port of the O2 sensor.

The probe assembly should be replaced if any part of it is cracked, or if liquid or other 
contaminate is visible inside the tubing. Minor kinks in the tubing are generally not a 
problem, but severe kinks can impede the sample flow or can develop a crack causing leaks. 
To replace, unscrew the black plastic screw from the fitting on the analyzer case. Inspect the 
interior of the fitting for any dirt or particles, which might cause a bad seal.  Screw in the 
plastic screw for the new probe.  The threads are easily stripped, so make sure that the screw 
is properly aligned before tightening. Turn until finger tight only. Do not use a tool to tighten, 
because this can strip the threads. 

   BATTERY OPERATION AND RE-CHARGING 

The Ni-Cd batteries in the 901 require several charge-discharge cycles before reaching full 
capacity. During the first 40-50 hours of use, the battery charge will not be optimum but will 
improve as the unit is used.  To properly condition the batteries, it is important during this 
initial period of use to run the unit in the battery mode until the batteries need to be recharged, 
as indicated by the LO BATT light. 

If left idle for 30 minutes, the unit will shut itself off to conserve power. When the battery is 
low, the LOW BATT indicator will be lit, indicating that the batteries need to be re-charged. 
The battery cover can be removed if needed by sliding it toward the bottom of the instrument.  

The unit will run 8-10 hours before needing recharging. To re-charge the batteries, plug the 
external charger module into a standard 115 Vac outlet and connect the plug from the charger 
to the jack located on the side of the 901. A full charge requires 8-10 hours of charging time. 
The unit will operate with four alkaline batteries, which are  non-rechargeable. However, 
these frequently leak and can corrode the terminals with long-term use. The unit is designed 
to operate with the batteries supplied. 

 Excessive charging can reduce the useful lifetime of Ni-Cad batteries. Unplug the unit from 
the charger when charging is complete. Do not operate the unit with the charger plugged in all 
the time. At least every 2-3 days, run the unit on battery power only. 

Coffee Laboratory 589 Rappahannock Drive White Stone Va 22578 TEL (804) 435-5522 

www.coffeelabequipment.com

Содержание 901

Страница 1: ...t internal sampling pump one main circuit board battery compartment NiCad rechargeable batteries battery charger module and sampling probe The sample probe is tipped with a particulate filter and luer fit syringe needle with side port holes to prevent plugging Sample is drawn through the probe and tubing and then to the oxygen sensor when the pump is turned on The pump is electronically timed to d...

Страница 2: ...the unit These batteries can overheat or crack if an extrnal voltage is applied and can melt the battery compartment The unit is fuse protected to prevent this If the unit will not operate under battery power alone but does operate with the charger plugged in the fuse is probably blown CALIBRATION SETTING Accurate SPAN calibration of the unit can be checked at any time by sampling room air which s...

Страница 3: ...pump if there are no leaks With this positive pressure air cannot leak into the system with any leaks which may be present If the reading goes lower it indicates a leak somewhere which should be corrected PUMP TIME The pump run time can be increased or decreased as needed by adjusting the PUMP TIME potentiomenter on the side of the unit Be careful not to set the time too high with samples that hav...

Страница 4: ...ool to tighten because this can strip the threads BATTERY OPERATION AND RE CHARGING The Ni Cd batteries in the 901 require several charge discharge cycles before reaching full capacity During the first 40 50 hours of use the battery charge will not be optimum but will improve as the unit is used To properly condition the batteries it is important during this initial period of use to run the unit i...

Страница 5: ...Ni Cd batteries designed to accept a quicker charge can also be used as long as they are AA size 1 2 1 4 volt and are of the nickel cadmium type Do not use other rechargeable types such as lithium AA alkaline non rechargeable batteries can be used in some situations but extreme caution must be exercised because plugging the unit into the charger with alkaline batteries installed can cause overheat...

Страница 6: ...idle Charger included Battery Life Eight hours before re charging Size 8 x 4 x 1 8 in 200 x 100 x 40 mm Weight 1 lb 450 gms OPERATING SUPPLIES 1 9001 Pkg of 200 foam seal pads 2 9002 Package of 1000 foam seal pads 3 9005 Particulate filter pkg of 5 4 9028 Particulate filter pkg of 25 5 9003 Side Port Needles pkg of 3 6 9029 Side Port Needles pkg of 15 7 9007 Sample probe assembly 8 9011 Battery Ch...

Страница 7: ...filter provided which fits onto the end of the sampling probe Avoid direct sampling of pressurized gas cylinders High pressure can cause damage to the sensor To test the content of a cylinder flush the sample into a plastic bag and sample from the bag Coffee Laboratory 589 Rappahannock Drive White Stone Va 22578 TEL 804 435 5522 www coffeelabequipment com ...

Отзывы: