Technical Brief

SLS-WAVE

qj

 Point of 

Administration Scanner

Overview

The Codonics SLS-WAVE Point of Administration 

Scanner is used in conjunction with the Safe Label 

System

®

 (SLS) Point of Care Station (PCS) and integrates 

with Cerner SurgiNet Anesthesia Information 

Management System (AIMS). The SLS and SLS-WAVE 

are used together in the operating room to improve 

safety and provide accurate, electronic documentation at 

the time of preparation and administration. 

Hardware Setup - SLS-WAVE/AIMS

1. Determine the location where the SLS-WAVE will be 

installed. For example, below an AIMS monitor or 
connected to an IV pole.

2. By default, the SLS-WAVE is shipped as if you are going 

to connect it to the left-side of the AIMS monitor or to an 

IV pole (see section 

Connecting to an IV pole

). 

SLS-WAVE Hardware Unpacking 
1. Remove all the components from the box. It should 

include:

SLS-WAVE scanner (1)

Back mounting bracket assembly (1)

Clamp mounting assembly (1)

Note: An open USB port is required to connect SLS-

WAVE into the computer where the AIMS application is 

operated.

Содержание SLS-WAVE

Страница 1: ...f preparation and administration Hardware Setup SLS WAVE AIMS 1 Determine the location where the SLS WAVE will be installed For example below an AIMS monitor or connected to an IV pole 2 By default th...

Страница 2: ...IV pole then go to section Connecting to an IV pole below If connecting to the right side of the AIMS monitor then go to the section Connecting to the right side of an AIMS monitor Connecting to the l...

Страница 3: ...r by rotating the platform thumbwheel as shown below 5 SLS WAVE mounted to the left side of the monitor 6 Insert the SLS WAVE USB into the AIMS computer WARNING The scanner should be pointing down and...

Страница 4: ...unting plate using a T 15 driver not included as shown below and move it to the other side of the mounting plate Using the T 15 driver screw the thick bracket back into the mounting plate 3 Remove the...

Страница 5: ...as shown below 8 The SLS WAVE is now mounted to the right side of the monitor 9 Insert the SLS WAVE USB into the AIMS computer WARNING The scanner should be pointing down and positioned to keep the sc...

Страница 6: ...tached to the IV pole by rotating the platform thumbwheel as shown below 5 The SLS WAVE is now mounted to the IV pole 6 Insert the SLS WAVE USB into the AIMS computer Configuring SLS WAVE with Cerner...

Страница 7: ...19 Fax Codonics Limited KK AQUACITY 9F 4 16 23 Shibaura Minato ku Tokyo 108 0023 JAPAN 81 3 5730 2297 81 3 5730 2295 Fax 3 To configure SLS WAVE for Cerner scan the following barcode below It will con...

Отзывы: