Setting up the BioXp™ system

Leave the BioXp™ system powered on

We recommend leaving the instrument on continuously. Only turn the instrument off when necessary, such as when moving the 

instrument to another location. If the instrument 

must

 be turned off, toggle the power switch to the 

Off

 position. 

Ethernet setup

The BioXp™ system requires an internet connection. Prior to connecting the instrument to the internet, be sure to test the 

connection. Connect the provided RJ45 ethernet cable to the ethernet port — located on the side of the unit — to an available 

ethernet wall jack.

In order for the BioXp™ system to communicate with the Codex DNA server, you must open the 80/443 port through 

your network. In addition, the BioXp™ system will be making an outbound SSL connection (HTTPS/SSL) to logmein.com, 

codexdna.com and drive.google.com. Unblock or whitelist these sites.

Start the BioXp™ system

1.  Ensure that all packaging materials have been removed, including the 

shipping screw located under the pipette head.

2.  Turn the instrument on using the power switch located on the side of the 

instrument.

3.  Close the instrument door. The BioXp™ system will initiate a self-check 

process to verify the correct functionality of motion control and other 

internal systems of the instrument. If the instrument is unable to perform a 

successful self-check it will alert the user to the problem and will prevent 

the instrument from operating until the problem has been addressed. The 

self-check process has been carefully developed to ensure that the BioXp™ 

system is operating within optimal calibration ranges and that all systems are 

functioning correctly before initiating a run.

 

Configure the BioXp™ system

1.  Press 

Configuration

Insert new job materials and

close door to start the next job.

Or change the configuration.

BioXp™ system ready

Configuration

2.  Press 

Network

Configuration

Date and time

Service

Network

Language

Done

Test the network connection

Press 

Test connection

If the system displays 

Network connection working normally

the BioXp™ system is ready for use. If you are ready to load 

the instrument and begin a run, refer to the appropriate deck 

loading instructions for your system run.

Network

Cancel

IP address

10.32.36.110

Subnet mask

10.32.36.110

Default gateway

10.32.36.110

Preferred DNS server

MAC address

10.32.36.110

MAC address

 

Use DHCP

Network connection
working normally

Test connection

OK

11:03:1D:12:8AF7

If the system displays 

No network detected

, contact 

Codex DNA customer service a

[email protected]

 for 

assistance in resolving the connection.

Network

Cancel

IP address

10.32.36.110

Subnet mask

10.32.36.110

Default gateway

10.32.36.110

Preferred DNS server

MAC address

10.32.36.110

MAC address

 

Use DHCP

No network detected

Test connection

OK

11:03:1D:12:8AF7

 

×

Содержание 43029

Страница 1: ...1 BioXp 3250 system User guide Part number 43029 Revision 1 1 Effective 11DEC2020...

Страница 2: ...IONS EXTENSIONS OR ADDITIONS TO THIS WARRANTY e You shall indemnify defend and hold harmless Codex DNA Inc from any costs expenses damages or other losses arising out of i any warranty of greater scop...

Страница 3: ...standards Alternating current Waste Electrical and Electronic Equipment WEEE directive Do not dispose symbol For more information see Equipment disposal and WEEE compliance Consult instructions for u...

Страница 4: ...d to an extension cord or to a multiple portable socket Doing so may compromise shielding and or grounding This equipment should not be used adjacent to or stacked with other equipment If it becomes n...

Страница 5: ...tem communication 3 Fuse Do not use 4 Power switch Switch power to the instrument on or off Note Leave the instrument on continuously 5 Power inlet Power the instrument BioXp applications The BioXp 32...

Страница 6: ...ent is unable to perform a successful self check it will alert the user to the problem and will prevent the instrument from operating until the problem has been addressed The self check process has be...

Страница 7: ...s 2 Add a minimum of 12 mL freshly prepared 70 ethanol to the reusable ethanol reservoir a Load ethanol reservoir in the right most reservoir retainer position of the instrument deck Note Do not disca...

Страница 8: ...ep running to continue the run Are you certain you want to cancel this job You will not be able to restart it Attention Yes Cancel job No Keep running If you press YES Cancel job the instrument will d...

Страница 9: ...technical services Additional information is included in the following table Read job issue Open the door to retry If this persists call 855 474 4362 Attention Unlock door General troubleshooting Code...

Страница 10: ...instructions may cause harmful interference to radio or television reception However there is no guarantee that interference will not occur in a particular installation If this equipment does cause in...

Страница 11: ...power quality should be that of a typical commercial or laboratory environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for...

Отзывы: