Page 13
Operation and Maintenance
Issue: A
Revision: 4
Date: 03/11/20
Ref:100255
3.1.1
Full Zero Cal
Allows user to activate Full Zero Cal.
Use
or
to scroll through to full Zero Cal.
Press
to activate Full Zero Cal.
The Data valid light shall extinguish, and the user will be returned to the Zero Calibration Sub-Menu. If the user re-
enters the menu “Wait for FZC” shall be displayed. The user should leave the menu and wait for the data valid light
to resume.
3.1.2
Bias Ct Zero Cal
Allows user to activate Bias current Zero Calibration.
Use
or
to scroll through to Bias Current Zero calibration.
Press
to activate Bias Current Zero calibration.
3.1.3
FZC Ref Time
Allows users to setup Zero Calibration time. The following are the submenus.
Set Full Zero Cal.
•
Use
or
to scroll through to FZC Ref Time press
.