The calibration routine must be run during commissioning,
otherwise the instrument will not be able to calculate the true
level within the duct. DO NOT run the calibration routine unless
reasonable conditions exist in the duct. If it is not the initial
calibration, it is recommended that the Gain Factor is recorded
from the parameters option, before the calibration is run.
4.10. Calibration for Dust Measurement
For Opacity Monitors to provide a measure of the dust density, the relationship
between opacity and the dust density needs to be established. For a particular
measurement point in a particular flue, duct, or chimney, this can only be
accomplished accurately by iso-kinetic sampling over a defined period. Over the
same period, the opacity monitor will calculate the corresponding average
extinction, which is directly proportional to the dust density
The relationship between extinction and dust density may be described as the
‘Dust Factor’.
Dust Density = Fine Dust Density/Extinction
Where :
fine dust density
=
mg/m
3
of dust by
iso-kinetic sampling
extinction
=
average extinction
for the same period.
This Dust Factor is input and held within the instrument to convert extinction to
fine dust density. Some important points to note when calibrating the from iso-
kinetic sampling results are given below.
Take care to ensure that the normalisation parameters
temperature, oxygen, pressure & water vapour) are set to sensible
values, and establish to what standard conditions the sampling
results will be normalised to. Iso-kinetic results are automatically
brought to ambient temperature as the flue sample cools through
the sample line, and in general these are then normalised to
0
C
(273K). Normalisation to the other parameters is variable, and
depends largely upon whether they are being measured at the
same time as sampling.
It is reasonable to start with an estimate of the dust factor see the
next page and ensure that at least the correct temperature is held
within the instrument, entered via the keypad. If it is uncertain
Содержание EnergyTech 100 Series
Страница 4: ...CODEL OPS 141 2 Issue A Revision 1 Date 14 06 18 Doc i d 0141 6 050028...
Страница 11: ...CODEL OPS 141 3 Issue A Revision 1 Date 14 06 18 Doc i d 0118 6 050028 Figure 1 General Arrangement...
Страница 18: ...CODEL OPS 141 10 Issue A Revision 1 Date 14 06 18 Doc i d 0141 6 050028 Figure 6 Connection Schedule...
Страница 35: ...CODEL OPS 141 27 Issue A Revision 1 Date 14 06 18 Doc i d 0118 6 050028 Figure 9 Program Tree...
Страница 41: ...CODEL OPS 141 33 Issue A Revision 1 Date 14 06 18 Doc i d 0118 6 050028...