OWNERS MANUAL | Cobolt 06-01 Series | D0136-K DECEMBER 2018
201
26 | 52
Direct On/Off control
06-MLD
This feature is not available for 06-MLD.
06-DPL
The Direct On/Off control feature enables turning the laser On/Off using a 5 VDC signal. After having configured the
laser for Direct Input operation (factory set or by executing @cobasdr 1), the laser can only start-up when 5 VDC (max
12.5 VDC) is applied to pin 3 on the Molex connector with 0 VDC on pin 2 as reference. Shifting the signal to
0
VDC on
pin 3 will turn the laser off and put the laser in stand-by mode. This input only controls the on/off state of the laser and
cannot be used to modulate the power output. The remote interlock connection between pin 1 and 2 must also be
made as described above.
Molex connector on back side of laser head.
NOTE
This function is not available for CDRH compliant models and cannot be used with key control box.
!