Using the Monitor

 

 

Power and Charging

 

 

The first time you use the monitor, use the wall adapter or a 
powered PC port and the USB cable to plug it in. 
The monitor has a built-in rechargeable battery. When the screen 
displays low battery, it should be recharged. The monitor can be 
used while charging. 
 

Turning the Monitor ON / OFF

 

 

Press the power button  to  turn the power on  or off.  When 
powered on, the display will show a 30 second count down. 
After the countdown is complete, you can change settings. 
 

Setting the CO2 Alarm Level, Date and Clock

 

 

Press the MODE button for two seconds to begin. The first 
setting is the CO2 alarm level. Press the left and right arrow 
buttons to select the digit to change. The selected digit will 
flash. Press the up and down arrows to change it. 

Press the MODE button again for 2 seconds to move to the 
date and clock values. Again, use the MODE button to 
change between values and the up and down arrows to 
change them. 

When complete, press the MODE button to save 

your changes and return to the normal display. The CO2 Alarm 

Level, Date and Clock can also be set using the software. See 

instructions at the end of this manual. 
 

Using the CO2 Alarm  

The CO2 alarm can warn you if carbon dioxide levels are too 
high. If the CO2 level goes over the level you set, a buzzer 
will sound. Press the left arrow button to turn the buzzer on 
or off.

 

The factory default carbon dioxide concentration alarm value 
is 2,000ppm. See below for recommended CO2 levels 

 

www.CO2Meter.com

 

www.CO2Meter.com 

6

 

Содержание XT-10

Страница 1: ...tions 2 Reading the Display 3 Using the Monitor 5 Carbon Dioxide Levels and Guidelines 7 Specifications 8 Software Installation 9 Software Application 12 Data Logging 13 Automatic Baseline Calibration 18 Manual Calibration 19 Contact Us 20 Carbon Dioxide Monitor Instruction Manual ...

Страница 2: ...ion Package Contents Monitor 110 220VAC to 5V DC Power Adapter USB Cable Instruction Manual Before You Start 1 This monitor contains sensors and microprocessors that use precision electronics It must be kept away from water heat fire combustible liquids and strong electromagnetic interference to prevent damage 2 For accurate results do not block the air sampling inlets Avoid strong or hot winds bl...

Страница 3: ...nctions Power Button turns the power ON OFF C F Up Arrow Button change between degrees Fahrenheit and Centigrade When MODE pressed used as up arrow or to increase the value 2 www CO2Meter com www CO2Meter com 2 ...

Страница 4: ... arrow or to decrease the value MODE MODE Button hold down for 2 seconds to enter SETTINGS Reading the Display 1 Carbon dioxide display area Battery level indicator PPM Concentration units parts per million CO2 Chemical symbol for carbon dioxide MON month MON DAY MIN SEC REC YEAR RH HOUR 2 3 2 Temperature display area degree Centigrade degree Fahrenheit YEAR year HOUR hour Sound alarm enabled 3 Hu...

Страница 5: ...arrow buttons to select the digit to change The selected digit will flash Press the up and down arrows to change it Press the MODE button again for 2 seconds to move to the date and clock values Again use the MODE button to change between values and the up and down arrows to change them When complete press the MODE button to save your changes and return to the normal display The CO2 Alarm Level Da...

Страница 6: ...s of concentration and increased likelihood of spreading respiratory viruses like colds etc 5 000ppm OSHA and NIOSH first threshold for safety maximum allowed concentration within a 8 hour working period Specifications Carbon Dioxide Range 0 9999PPM Accuracy 70ppm 3 of reading 0 5 000ppm Resolution 1ppm Response time 1 second Temperature Measurement 10 to 70 14 158 Temperature Resolution 0 1 Tempe...

Страница 7: ...ile to your Windows PC Double click the file to start the installation Click Next to begin Accept the default installation folder on your hard drive or click Browse to create a new one Click Next after confirmation and click Next again to start the installation 9 www CO2Meter com www CO2Meter com 10 ...

Страница 8: ... cable to connect the computer and the CO2 monitor Click to connect The following prompt indicates successful connection Click to disconnect Setting the Date Clock After connecting to the computer click Set Time to automatically set the clock The monitor will sync with the computer s built in clock 11 www CO2Meter com www CO2Meter com 12 ...

Страница 9: ...Click the Immediately button to tell the software you want to log data from the software then click OK to begin data logging While logging REC will flash on the display The flashing frequency is the same as the recording interval you set in the Param screen When you are done data logging click Stop Recording The data log cannot be observed until it is downloaded to your PC To download click Data D...

Страница 10: ...a room start recording then come back later and retrieve the monitor and download the data This is useful for measuring air quality in a room over time Start by connecting the monitor to a PC and running the software After connecting click the Set Param button to enter the parameter setting interface as shown in the figure below 15 www CO2Meter com www CO2Meter com 16 ...

Страница 11: ...previously logged data in the monitor s memory is cleared Therefore you should get in the habit of saving your data files every time you create one Automatic Baseline Calibration Automatic baseline calibration is useful in areas that are regularly unoccupied and CO2 level return to 400ppm every few days ABC is turned off by default To enable ABC Step 1 Turn the power off Step 2 While pressing the ...

Страница 12: ...onitor so that your breath does not influence the manual calibration Manual calibration will take between 5 10 minutes to complete Step 6 After calibration is complete read the display 1 means manual calibration was a success 0 means it failed usually because the CO2 level was not stable during calibration If it fails retry the manual calibration again If it fails repeatedly contact us for support...

Отзывы: