background image

 

OPERATION MANUAL 

Desktop CO2, Temperature & 

Relative Humidity Monitor   

TIM‐12 

 

 

 

INTRODUCTION 

Thank you for purchasing the TIM12 desktop 
CO2 monitor. It measures carbon dioxide levels, 
air temperature and humidity. This makes it 
ideal for indoor air quality (IAQ) diagnosis and 
HVAC system performance verification. 
 
Poor indoor air quality is considered unhealthy. 
It can lead to drowsiness, inability to 
concentrate, and even illness. 
 
IAQ monitoring of CO2 levels is widely used in 
public areas such as offices, classrooms, 
factories, hospitals and hotels. It is also 
recommended for industrial hygiene in some 
countries. (See appendix). 

Product Features: 

● Large display shows CO2 level, temperature, 
    humidity, date and time 
● 15 degree tilt angle makes the display easy 
    to see and read 
● Indoor air quality level indication: 
    Good, Normal, and Poor 
● Stable NDIR sensor for CO2 detection 
● Visible and audible CO2 warning alarm 
● ABC (Automatic Baseline Calibration) 
    or manual CO2 calibration 
● Maximum or minimum CO2 level recall 
● Sampling rate from 1 sec. to 4:59:59. 
 

MATERIALS SUPPLIED 

This package contains: 

● Meter 

● Power Supply (5V +/‐ 10%, >= 0.5A) 

● USB cable and software CO 

● Manual 

POWER SUPPLY 

The meter is powered by an AC adaptor 

(5V/0.5A output). 

 

LCD DISPLAY 

 
 
 
 
 
 

 

Symbols 

ppm                  CO2 level in parts‐per‐million 

GOOD                CO2 is in good level 

NORMAL            CO2 is in normal level 

POOR                CO2 is in poor level 

Air Temp.          Air temperature 

Humidity %        Unit of air relative humidity 

o

E (C/F)              Celsius/Fahrenheit of temp. 

MAX/MIN          Maximum/Minimum readings 

                          CO2 alarm icon 

 

NOTE:   

TWA / STEL / VENT.R / are vain icons on this 

model. 

 

 

KEYPAD 

SET 

‐Enter setup mode 

‐Save and finish settings 

ESC

‐Exit setup page / mode 

      ‐Terminate calibration/data logging 

RESET 

‐ Clear MAX/MIN values 

LOG

‐ Select mode, increase value at setup 

‐ Start data logging 

MIN/MAX

▼ 

‐Activate MAX, MIN function 

‐Select mode or decrease value 

during setup 

SET

+

+

MIN/MAX

▼ 

‐ CO2 calibration mode 

 

OPERATION 

POWER ON/OFF 

Plug the wall adaptor into an appropriate 

receptacle. The meter will turn on automatically 

after a short beep. The lit LED indicates full 

power. If the voltage is too high or low, “bAt” 

will display on the LCD and the LED will flash. 

See page 14 for trouble shooting. 

 

During normal operation, the LCD will display 

the current CO2 level in ppm, air quality level, 

air temperature, humidity, and the current date 

and time (Fig.1). 

 

 

 

 

 

 

TAKING MEASUREMENTS 

The meter starts taking measurements 

immediately after it is turned on and updates 

the readings every second. In the event the 

operating environment is quickly changed (for 

example, from high to low temperature), it will 

take approximately 2 minutes for the CO2 and 

temperature to adjust and 10 minutes for 

humidity to stabilize. 

NOTE:

 Do not hold the meter close to your face 

as your breath will affect the CO2 level. 

 

MIN/MAX 

While the meter is powered on, press the 

MIN/MAX

“ button to see the minimum and 

maximum CO2 level, temperature, and 

humidity in turn. Each press of “

MIN/MAX

“ will 

display the MIN and MAX values in order, then 

return the meter to normal mode (Fig.2). 

 

During normal operation or while viewing 

MIN/MAX levels, press and hold the 

RESET

“ key for more than 1 second to clear the 

stored minimum and maximum values. 

 

 

 

 

 

 

 

 

 

INTERNAL DATA LOGGING 

Press “LOG

” for 2 seconds under normal mode 

to start logging. The green LED blinks to indicate 

the logging status and the LCD shows the real 

time CO2 value and “rEC”. Lower displays are 

the real time temperature, humidity and clock. 

 

To terminate data logging, press “ESC” for 2 

seconds, The LED light will stop blinking and the 

display will show “End” and CO2 reading 

interchangeably. Hold down “ESC” for 2 

seconds again, and it will return to normal 

measurement mode.   

 

The meter can save CO2, temperature and 

humidity levels over time in internal memory. 

The memory capacity is 15,999 points. The 

sampling rate can be from 1 second to 4 hours 

59 minutes and 59 seconds (See SETUP). 

Minimum and Maximum recall is still available 

during logging. 

 

 

 

 

 

HIGH CO2 ALARM 

Содержание TIM-12

Страница 1: ...n mode OPERATION POWER ON OFF Plug the wall adaptor into an appropriate receptacle The meter will turn on automatically after a short beep The lit LED indicates full power If the voltage is too high or low bAt will display on the LCD and the LED will flash See page 14 for trouble shooting During normal operation the LCD will display the current CO2 level in ppm air quality level air temperature hu...

Страница 2: ...X to decrease the value The POOR level can be set between 1 000 to 5 000ppm Press SET to confirm P1 3 and continue to P1 4 Fig 8 or press ESC without saving to return to P1 0 P1 4 SET ALARM ON OFF P1 4 is used to set alarm on and off When entering P1 4 the default On will be blinking on the LCD Fig 8 Press or MIN MAX to switch the alarm to Off Then press SET to confirm P1 4 and return to P1 0 Fig ...

Страница 3: ...n is enabled by default from the factory For ABC to work properly the meter must continuously monitor CO2 levels for at least 7 5 days power on During that time it will record the lowest CO2 level and set that as the 400 ppm baseline for future readings Therefore it is critical when using ABC calibration that the meter be exposed to fresh air at least once every 7 5 days If this is not possible th...

Страница 4: ...ke sure you secure the unit inside the package properly to prevent any damage during transit that could void your device s warranty Finally please ship your device to the address shown under the Contact Us section below CO2Meter will not under any circumstances be responsible for your shipment expenses and no refund will be issued for shipping charges necessary for you to ship the unit to us Liabi...

Отзывы: