background image

 

Calibration

 

Procedure:

 

RAD

 

CO2

 

Safety

 

Alarms

 

This

 

is

 

a

 

more

 

detailed

 

explanation

 

of

 

the

 

calibration

 

procedure

 

found

 

in

 

the

 

manual.

 

Check

 

your

 

state

 

or

 

local

 

code

 

for

 

calibration

 

schedule

 

requirements.

                     

At

 

a

 

minimum,

 

you

 

should

 

calibrate

 

at

 

least

 

once

 

a

 

year.

 

There

 

are

 

2

 

different

 

calibration

 

procedures

 

depending

 

on

 

the

 

altitude

 

at

 

which

 

the

 

device

 

is

 

to

 

be

 

used:

 

1.

 

0

 ‐ 

3,000

 

ft.

 

(900m):

 

Perform

 

Zero

 

Calibration

 

and

 

Span

 

Calibration.

 

2.

 

3,000+

 

ft.:

 

Perform

 

Zero

 

Calibration

 

only.

 

 

Before

 

You

 

Begin

 

1.

 

For

 

Zero

 

Calibration,

 

you

 

will

 

need

 

a

 

gas

 

cylinder

 

of

 

pure

 

nitrogen

 

(0

 

ppm

 

CO2).

 

For

 

Span

 

Calibration,

 

you

 

will

 

need

 

a

 

2

nd

 

cylinder

 

of

 

4%

 

CO2.

  

2.

 

Temporarily

 

cover

 

the

 

buzzer

 

hole

 

in

 

the

 

upper

 

right

hand

 

corner

 

of

 

the

 

SEU

 

(sensor

 

unit)

 

and

 

RDU

 

(remote

 

display

 

unit)

 

with

 

a

 

small

 

piece

 

of

 

masking

 

tape

 

to

 

minimize

 

the

 

alarm

 

sound

 

during

 

calibration.

 

3.

 

If

 

calibration

 

is

 

to

 

be

 

performed

 

in

situ,

 

first

 

inform

 

building

 

occupants

 

that

 

you

 

will

 

be

 

testing

 

the

 

CO2

 

alarm

 

for

 

several

 

minutes.

 

4.

 

Attach

 

included

 

tubing

 

between

 

the

 

gas

 

cylinders

 

and

 

the

 

SEU

 

calibration

 

port

 

as

 

follows:

 

a.

 

RAD

0102SX2

 

 

Connect

 

the

 

tubing

 

to

 

the

 

supplied

 

plug

 

and

 

secure

 

it

 

over

 

the

 

air

 

entry

 

/

 

calibration

 

port

 

on

 

the

 

front

 

of

 

the

 

SEU.

  

b.

 

RAD

0102/RAD

0200

 

 

Connect

 

the

 

tubing

 

to

 

the

 

calibration

 

port

 

on

 

the

 

bottom

 

of

 

the

 

SEU.

 

 

    

A:

 

RAD

0102SX2

  

    

B:

 

RAD

0102/RAD

0200

 

Отзывы: