Background calibration 

Calibrate the sensor while the reading deviates from the actual gas 

concentration. 

Change calibration and ABC setting 

parameter. 
Automatic background calibration 

Operation Instructions 

1.

Power on and Measure:

Connect the USB cable to a PC or a USB to wall adapter to power on the device. After powered on, the bars on the LCD will decrease by degrees

during warm-up. The device will start to measure CO

concentration, ambient temperature and relative humidity after warm-up.

2.

Backlight Display Mode:

"

Auto

" auto backlight is the default mode. Press Mode key (

2

 

) to display the current 

back-light

 mode and then press Mode key (

2

 

) to select 

setting modes in sequence: 

Auto

 

 

AL 

 

on 

 

OFF

. After selecting the mode, it is the confirmation mode. You do not need to press any key to

confirm

 

it.

2.1 

Auto

: The backlight color varies with the CO

2

 concentration. <800ppm is green, 

good status

;

 

9

50

ppm~1,400ppm is yellow, 

okay status,

>1400ppm is red, 

poor status – requires ventilation

.

2.2 

AL

: The alarm level is exceeded and the backlight is displayed in red. The default alarm level is 1,400ppm. Please follow the alarm level setting

method to adjust the alarm level. 

2.3 

on

: The LCD backlight is always on. It is only LCD backlight, without any prompt function. The backlight color depends on the color you select. 

(please see the setting of 

COL

). 

2.4

 OFF

: Turn off all backlight functions. 

3. Setting:

Press and hold Mode key (

2

) to enter the settings. And then press Mode key (

2

) to select setting modes in sequence: 

COL

 (LCD backlight color)

AL 

(alarm level setting) 

 

Real-time 

 

ECO 

(energy-saving mode). Press Select key (

3

) to adjust the parameters.

3.1 

COL

: The default LCD backlight color is white. A total of 7 colors are optional. The color order is white 

 

turquoise 

 

blue 

 

purple 

 

red 

 

yellow 

 

green. (Figure 1) 

3.2 

AL

: The default alarm setting is 1,400ppm. The alarm value can be adjusted from 400ppm to 3,000ppm (the interval is 100ppm). (Figure 2) 

3.3 

Real-time

:   

3.3.1 Set “12 Hour (AM/PM)/24 Hour” by pressing Select key (

3

).

 

3.3.2 The default real-time setting is 0:00(24 Hour Clock). The setting sequence is hour 

 

minute. After setting "hour", press Mode key (

2

) to 

set "minute" (the interval is 1 hour / 1 minute) (Figure 3). Press Select key (

3

) to adjust the parameters. Press and hold Select key (

3

to quickly set the parameters.   

3.4 

ECO

: The default ECO setting is “OFF’. Press Select key (

3

) to select

 

on

’.

 

After ECO mode is turned on, the LCD backlight will be dark 

during the period from 22:00 pm to 6:00am for saving power. (Figure 4) 

(Figure 

1)

(Figure 2)

 

(Figure 3)

 

(Figure 4)

 

Note: During the setting process, the device will return to the main screen after 20 sec of idle time.

 

4.

°C/°F

 

switch: 

P

ress and hold 

Select key (

+

)

 to switch °C/°F

.

5.

Calibration

1.

Plug in and power your device outside.

***Note: Do not breathe directly on your device during the calibration process. 

***Note: Do not place the device in direct sunlight.

2.

Allow your device to sit outside for 3 - 5 minutes to allow CO2 levels to adjust appropriately.

3.

Press and hold the Mode key (

MODE

) and then press Select key (

+

) together. When the CO2 calibration value flashes on the LCD, it means 

that the calibration mode has been entered.