BT61i   pH   Sensor   User’s   Guide   

|

   

3

   

Calibration      

The    CMA    pH    sensor    is    supplied    calibrated.    The    output    of    the    sensor    is    linear    with   
respect   to   pH   value.   The   supplied   calibration   function   is:         

   

pH   =   -­‐4.04   *   V

out

   (V)+   13.68   

   

The    Coach    program    allows    selecting    the    calibration    supplied    by    the    sensor    memory   
(EEPROM)   or   the   calibration   stored   in   the   Coach   Sensor   Library.   For   better   accuracy   a   
new   user   calibration   can   be   performed   in   the   Coach   program.   In   order   to   calibrate   a   
pH   Sensor,   you   should   have   a   supply   of   pH   buffer   solutions   that   cover   the   range   of   pH   
values   you   will   be   measuring.   To   perform   a   two-­‐point   calibration:   

 

Rinse   the   tip   of   the   electrode   in   distilled   water.      

 

Place    the    electrode    into    one    of    the    buffer    solutions.    When    the    voltage    reading   
stabilizes,   enter   the   pH   value.   

 

For   the   next   calibration   point,   rinse   the   electrode   and   place   it   into   a   second   buffer   
solution.      

 

When   the   voltage   reading   stabilizes,   enter   the   second   pH   value.      

 

Test   your   calibration   in   different   known   pH   buffer   solutions.   

Suggested   experiments      

The    pH    sensor    can    be    used    for    various    experiments    in   
biology,   chemistry   and   environmental   science   such   as:      

 

Measuring   pH   values   of   different   acids   and   bases,   

 

Acid-­‐base   titration   experiments,   

 

Monitoring   pH   during   chemical   reactions,   

 

Investigating   of   water   quality   in   streams   and   lakes.   

   

Practical   information   

 

Do   not   place   the   electrode   in   base   solution   (pH   >   10)   for   longer   than   few   hours.   
This   can   affect   the   glass   of   the   electrode.      

 

The    pH    electrode    has    a    limited    operational    life    and    can    be    ordered    separately.      
During   a   measurement   the   end   of   the   plastic   tube   of   the   electrode   has   to   be   held   
approximately   1   cm   in   the   liquid.      