background image

BT57i   ORP   sensor   User’s   Guide   

|   3

   

As   a   result,   you   do   not   have   to   perform   a   new   calibration,   just   use   the   calibration   
supplied    by    the    sensor    memory    (EEPROM)    or    the    calibration    stored    in    the    Coach   
Sensor   Library.      
If    you    are    doing    water    quality    testing    or    performing    a    chemistry    experiment    that   
requires    a    very    accurate    calibration,    you    will    need    to    obtain    two    commercial    ORP   
standards.   Using   these   standards,   perform   the   following   2-­‐point   calibration:   

 

For   the   first   calibration   point,   rinse   the   tip   of   the   electrode   with   distilled   water,   
and   place   the   electrode   into   the   first   standard.   When   the   voltage   reading   
displayed   by   the   data-­‐collection   program   stabilizes,   enter   the   ORP   value   (mV)   of   
the   first   ORP   standard.   

 

For   the   second   calibration   point,   remove   the   electrode   from   the   first   standard,   
rinse   it   with   distilled   water,   and   place   it   into   the   second   standard.   When   the   
displayed   voltage   reading   stabilizes,   enter   the   ORP   value   (mV)   of   the   second   
standard.      

 

Rinse   the   electrode   with   distilled   water   and   place   it   into   the   sample   to   be   
measured.   You   are   now   ready   to   take   measurements   with   the   calibrated   ORP   
sensor.   

Suggested   experiments      

Redox    reactions    control    the    behavior    of    many    chemical    constituents    in    drinking   
water,   wastewater   and   aquatic   environments.   The   reactivity   and   solubility   of   critical   
elements   in   living   systems   are   strongly   dependent   on   redox   conditions.   ORP   values   
are   used   much   like   pH   values   to   determine   water   quality.   For   example   for   swimming   
pools   at   a   normal   pH   value   between   7.2   and   7.6,   the   ORP   value   must   be   kept   above   
700    mV    to    kill    unwanted    organisms.        In    contrast,    in    order    to    support    life,    natural   
waters    need    a    much    lower    ORP    value.    Generally    ORP    values    above    400    mV    are   
harmful   to   aquatic   life.      
The   ORP   sensor   can   also   be   used   for   redox   titrations   to   determine   the   equivalence   
point   in   an   oxidation-­‐reduction   reaction.      

Отзывы: