BT32i   Differential   Voltage   sensor   User’s   Guide      

Short   description   

The   Differential   Voltage   sensor   BT32i   is   designed   for   measuring   voltages   in   the   range   
between    -­‐500    and    500    mV.    The    sensor    is    a    signal    amplifier    with    a    wide    frequency   
range,   which   can   be   used   to   measure   small   voltages   in   AC   and   DC   circuits.      
The    sensor    has    differential    inputs,    which    means    that    measurements    can    be    done   
directly   across   circuit   elements   without   the   constraints   of   common   grounding.   It   can   
be    used    to    measure    positive,    as    well    as    negative    potentials.    It    has    two    banana    (4-­‐
mm)   plugs   for   easy   connection.   

   

The    sensor    should    be    connected    parallel    to    a    circuit    element.    It    measures    the   
potential   difference   between   the   V

+

-­‐   red   plug   and   the   V

-­‐

   -­‐   black   plug.   The   measured   

voltage   goes   through   an   amplifier   unit   and   the   output   of   the   sensor   is   adjusted   to   
the   range   of   

±

7.5V.   The   sensor   is   provided   with   over-­‐voltage   protection   and   voltages   

up   to   

±

50V   (related   to   ground)   will   not   damage   the   sensor.   It   never   can   be   used   for   

higher   voltages.   

   

The   Differential   Voltage   sensor   can   be   directly   connected   to   analog   BT   inputs   of   the   
CMA   interfaces.   The   sensor   cable   BT   -­‐   IEEE1394   needed   to   connect   the   sensor   to   an   
interface   is   not   supplied   with   the   sensor   and   has   to   be   purchased   separately   (CMA   
Article   BTsc_1).      

Sensor   recognition   

The   Differential   Voltage   sensor   has   a   memory   chip   (EEPROM)   with   information   about   
the    sensor:    its    name,    measured    quantity,    unit    and    calibration.    Through    a    simple   
protocol    this    information    is    read    by    the    CMA    interfaces    and    the    sensor    is   
automatically   recognized   when   it   is   connected   to   these   interfaces.      If   your   Voltage   
sensor    is    not    automatically    detected    by    an    interface    you    have    to    manually    set    up   
your   sensor   by   selecting   it   from   the   Coach   Sensor   Library.      

Calibration      

The    CMA    Current    sensor    BT32i    is    supplied    calibrated.    The    output    of    the    sensor    is   
linear   with   respect   to   the   input   voltage.      The   supplied   calibration   function   is:   

   

V

in

   (mV)   =   78.125   *   V

out

   (V)   –   0.47   where   V

in

   =   V

+

   -­‐   V

-­‐

   

   

The   Coach   program   allows   selecting   the   calibration   supplied   by   the   sensor   memory   
(EEPROM)   or   the   calibration   stored   in   the   Coach   Sensor   Library.   For   better   accuracy   
the   pre-­‐defined   calibration   can   be   shifted.      
For   even   more   accurate   measurements   a   new   user   calibration   (a   standard,   simple   2-­‐
point   calibration)   can   be   performed   in   Coach   using   known   voltages.      
   