Technical data 

Input current range  

± 500 mA 

Output voltage range 

± 7 V 

Sensitivity 12.8 

V/A 

Resolution using 12 bits A/D converter  0.38 mA  

Calibration function  

I

in

 (mA) = 78.125 * V

out

 (V) – 0.47  

Probe resistance

2

  

typical 1.3 

 

Input impedance to ground 

each terminal 400 k

  

Input offset current error 

typical  

 0.8 mA  

Common mode input voltage error 

typical 0.15 mA/V  (0 – 500 Hz)      

Non-linearity 

<  0.001 % 

Slew rate  

3 V/

s (maximum output voltage variation) 

Bandwidth (small signal) 

120 kHz (-3dB) 

Maximum differential input voltage 
Maximum common-mode input voltage 

± 50 V (max. voltage between input terminals)  
± 50 V (max. voltage related to ground) 

Supply voltage / Supply current  

5 V DC / typical 23 mA 

Sensor information for Auto-ID and 
calibration 

256 byte serial EEPROM  

Connection 

Right-hand BT (British Telecom) connector  

 
Warranty:  

The 0222i Current sensor is warranted to be free from defects in materials and 
workmanship for a period of 12 months from the date of purchase provided that it 
has been used under normal laboratory conditions. This warranty does not apply if 
the sensor has been damaged by accident or misuse.  
 

 

Note:

 

This product is to be used for educational purposes only. It is not appropriate 

for industrial, medical, research, or commercial applications.    

 

 

 

 

 

Rev. 1/11/12

 

                                                

 

 

2

 

The probe resistance is a compound of a 0.4 

 shunt and a 0.9 

 multifuse. The multifuse protects the 

shunt against overload. The time to trip the multifuse to a high-resistant state is 0.1 sec. at 5 A. 