BT24i   CO

2

   Gas   Sensor   User’s   Guide   

|   

3

   

having   a   CO

2

   concentration   of   about   400   ppm.   While   still   outdoors   insert   the   sensor   with   the   

rubber   stopper   into   the   bottle.   You   can   now   take   the   bottle   and   the   sensor   to   the   location   where   

the   measurements   will   be   done.      

 

Connect   the   CO

2

   sensor   to   an   interface.   

 

Let   the   sensor   warm   up   by   collecting   data   for   at   least   90   seconds.      

 

When   the   readings   are   stabilized   use   a   paper   clip   or   a   ballpoint   to   press   down   the   calibration   
button.   After   about   30   seconds,   the   reading   should   stabilize   at   a   value   of   approximately   400   ppm   

(±40   ppm).      

 

If   the   reading   is   significantly   lower   or   higher   than   400   ppm,   simply   press   the   button   again   to   
repeat   the   process.      

Collecting   data      

1.

 

Connect   the   sensor   to   your   interface   and   let   it   warm   up   for   at   least   90   seconds   to   assure   a   stable   

voltage.   

2.

 

You   can   use   the   sampling   bottle   to   collect   air   samples.      

3.

 

Calibrate   your   sensor   by   using   its   calibration   button.   See   the   calibration   procedure   above.      

4.

 

Start   your   measurement.   The   sensor   takes   a   measurement   reading   about   every   5   s   (the   intensity   

of   IR   changes).   

5.

 

Even   though   the   sensor   responds   rather   quickly   to   changes   in   CO

2

   concentration,   remember   that   

gas   has   to   diffuse   through   the   holes   in   the   sensor   tube   before   any   changes   in   concentration   will   be   

detected.   Since   diffusion   of   gases   is   a   fairly   slow   process,   there   is   a   resulting   delay   in   the   readings.      

Practical   information   

 

The   sensor   requires   a   large   current   (120   mA).   We   recommend   that   you   use   the   power   adapter   for   
data-­‐loggers.   Only   one   CO

2

   sensor   can   be   used   at   a   time.      

 

The   sensor   cannot   take   readings   at   CO

2

   concentration   higher   than   5000   ppm.   Once   the   CO

2

   

concentration   reaches   this   level   the   computer   will   continue   to   display   a   reading   of   5000   ppm,   until   

the   actual   level   drops   below   5000   ppm   again.      

 

The   sensor   is   designed   to   operate   between   20°C   and   30°C.   The   sensor   can   be   used   outside   of   this   
temperature   range;   however   there   will   be   a   loss   in   accuracy   of   readings,   even   if   the   1-­‐point   

calibration   at   the   lower   level   or   higher   temperature   is   done.   Allow   enough   time   for   the   sensor   to   

stabilize   at   the   desired   operating   temperatures.      

Suggested   experiments      

 

Measuring   CO

2

   levels   (respiration)   from   small   animals   and   insects.   

 

Monitoring   CO

2   

changes   in   a   plant   terrarium   during   photorespiration   and   photosynthesis   in   

light/dark   (combination   with   the   light   sensor).   

 

Measuring   CO

2

   levels   during   cellular   respiration   of   peas   or   beans.   

 

Monitoring   production   of   CO

2

   during   chemical   reactions.   

 

Measuring   CO

2

   levels   in   classroom   

   

   