background image

2

Technical data

Shutdown procedure

  For normal isolation, the Closomat toilet can be 

turned off using the integrated switch. To carry out 

any servicing, the unit 

MUST

 be isolated at the 

mains feed.

Bathroom/shower room 

installations

•  

 The Closomat toilet must not be in direct line  

with the shower spray. Some form of partition 

between shower and the Closomat toilet is 

advisable to prevent any water spraying directly 

onto the equipment.

•  

 A shower should not be used in conjunction with 

the Closomat toilet i.e. under no circumstances 

must a user shower whilst seated on the        

Closomat toilet.

Water services (WRAS approved)

  

It is essential that the entering water 

is cold.

  We would recommend water supply isolated via 

isolation valve.

  Water supply can be from storage vessel or mains 

(‘Grey Water’ 

MUST NOT BE USED

).

  Boiler Capacity: 1.7 litres.

  Water rate:  min. 8 l/min.  

max. 0.8 MPa (8 bar)

  Cistern capacity: 6 litres.

Overflow

  The Closomat toilet has an internal overflow 

which discharges into the W.C. pan through the 

flush valve.

Soil connection

  Relevant soil connector to be sourced by installer.

  Soil connections suitable for Closomat toilet;      

S-Trap (vertical fall) wall to centre of pipe from 

70mm-230mm max. P-Trap (horizontal) floor to 

centre of pipe 200mm max.

  The Closomat toilet is suitable for all turned trap 

outlets, through purpose made removable side 

covers on lower level.

The Closomat toilet 

MUST

 

be installed by a competent or 

qualified person.

Explanation of symbols

  Safety instructions in this manual are identified by 

symbols. The safety instructions are introduced 

by signal words, which express the extent of the 

hazard/risk.

  In order to prevent accidents, personal injury, 

property damage and guarantee maximum user 

safety, follow the safety instructions without fail 

and exercise caution.

Description of equipment

  The Closomat toilet is an electrically operated 

shower (wash and dry) toilet designed to provide 

the operator/user with a hygienic solution to toileting.

 

Start-up procedure

  To enable a warm washing cycle operation 

the Closomat toilet must be switched on.             

Each Closomat toilet has its own isolation switch.  

To switch the unit on, locate the switch on the 

right hand side of the unit at the base of the 

cistern cover, through a Perspex screen.

  Once switched on, the water will reach core 

temperature within 5 minutes. It is advised that to 

maintain the core temperature in readiness for 

use, the unit be left switched on at all times.

Electrical services

  To avoid the risk of electric shock, this equipment 

MUST

 only be connected to a supply mains with 

protective earthing by a competent or qualified 

person.

Mains isolation

  All electrical appliances must be installed in 

accordance to the relevant regulations associated 

with the country it is within. To conduct any form 

of repairs or servicing it is recommended that       

a mains isolation device be located in the vicinity    

of the unit.

  220-240V single phase AC supply is required, 

(loading 4.5A-4.9A, maximum power 1300 watts).     

The Closomat toilet is supplied with a 3 core cable 

attached and 

MUST NOT

 be removed or replaced. 

A 10A fused spur is required for isolation.

  Spur outlet to be located in accordance with 

current I.E.E. Regulations (Current Edition).

  No modification of this unit is allowed other than 

by the manufacturer.

  

Closomat has IPX4 rating.

Содержание Palma Vita

Страница 1: ...Closomat Palma Vita installation user guide...

Страница 2: ...introduced by signal words which express the extent of the hazard risk In order to prevent accidents personal injury property damage and guarantee maximum user safety follow the safety instructions w...

Страница 3: ...ed within a typical shower toilet room Infill panel The infill panel specifications are printed on page 11 Classification Low Voltage Directive EN 60335 2 84 Cleaning information The use of bleach pro...

Страница 4: ...for standard pan connection fittings D Rear wall fixing Position 25mm to RH of the centreline 810mm from the floor R Rear floor fixing 2No Standards Installations Position on the floor 102mm from the...

Страница 5: ...er completely but do not remove it IMPORTANT Please retain all packaging until the installation is complete Protect all floor surfaces when handling the Closomat DO NOT slide the Closomat across the f...

Страница 6: ...ion the Closomat and place all four anchor blocks in place 15 Secure the Closomat to the floor using all four fixing points use suitable fixing for the application Adhesives are not allowed to be used...

Страница 7: ...rm to hold in this extended position arrow Run wash cycle to purge Once wash is constant remove tool and allow the douche arm to retract 2 Allow approximately 5 minutes for the water to heat in the bo...

Страница 8: ...bow pads 8 Replace the seat by locating the seat pivots into the seat hulls fixed to the pan press the seat firmly to secure its position 7 Push in the seat buffer on the cistern cover 4 Replace ciste...

Страница 9: ...it the side panel and screw covers on completion 1 Using the template provided locate mark and drill four holes for the plinth in the floor to best suit the intended fixings to be used for the applica...

Страница 10: ...armrest is moved Important Ensure Belville Washer is assembled correctly with the dome towards the bolt head Hole diameter 76mm minimum 78mm maximum Measure 367mm from rear of unit to centre of arm fi...

Страница 11: ...installation where pipe work inhibits the Closomat toilet from being located against the wall The infill panel is available in thickness sizes 25mm 50mm and 75mm The infill panel is 100mm taller than...

Страница 12: ...losomat co uk closomat co uk British manufacturing British design and development British customer support We reserve the right to make modifications as a result of ongoing adjustments and further dev...

Отзывы: