B. Status indica�on
Power status indicator: The icon of power
status has two different indicators, power-
on icon and power-off icon.
Door status indica�on: The icon of door status
has two different indicators, door-on icon and
door-off icon.
Notes:
When we are shoo�ng, we must make
sure that the door is in “closed” status.
Gel-cu�ng icon: The icon of gel-cu�ng status
has two different indicators, gel-cu�ng-on icon
and gel-cu�ng-off icon.
Warning:
1. Before cu�ng gel, please check safety items on
page P3.
2. Opera�on methods of the gel cu�ng, please refer
to P38 in user manual.
C. Light source control
Click icon to turn on the double-sided LED EPI white light.
This func�on is also used to adjust the sample posi�on, f-
ocus precisely and enter the preview func�on of the instr-
ument.
Notes:
The preview func�on of instrument need to set va-
rious parameters before you use this func�on. Please refer
to P27-P30 in this chapter.
Click icon alternately to power the Super Slim trans-
white/trans-blue LED transilluminator on or off.
White LED Transilluminator: apply for the Protein gel.
Blue LED Transilluminator: apply for the nucleic acid
gel with dyes such as Gel Green and Sybr Green.
Blue LED (white LED)Transilluminator Installa�on refers
to p13-p14 in user manual.
UV Transilluminator : Click icon alternately to power the
Super Slim trans-UV LED transilluminator on or off.
UV Transilluminator: apply for the nucleic acid gel dyed
with dyes such as EB or Gel Red.
D.Camera parameter Se�ng
Set aperture:
We use the seven digit numbers to cover adju-
stment of the aperture size. The larger aperture value (1-7)
is, the larger light transmission is and the brighter backgrou-
nd is because the customer is not expert of op�c filed.
Doing in this way, it can avoid customs confusing the generic
term, such as F1.4, F2, and F2.8.
In the le� picture: aperture value is 7, the stronger
captures light signal, and the background is brigh-
ter.
General parameter of aperture:
Under UV transmission, aperture value is 5-6.
Under blue light transmission, aperture value to 4-5.
On Off
26
27