
Models AC-30B and AC-30BHP Mains Frequency Spark Tester
AC-30B Instruction Manual
–
Page 48
The red electrode mounting plate, protective cover, and bead chain assemblies
are subject to damage and contamination that is not always visible. They should
be replaced if current leakage occurs.
Refer to the “Troubleshooting”
section for assistance with electrical problems.
Troubleshooting
CAUTION:
Troubleshooting is to be performed by qualified service personnel
only. Failure to follow the procedures in this manual may result in danger to
personnel and equipment damage.
Phantom Faults are being indicated.
1.
Lateral wire line vibration or water may be present. See “Installation” for
information on drying, centering and restraining the product in the
electrode.
2.
The high voltage mounting plate may be contaminated with dirt or
conductive material. Clean the mounting plate or replace.
3.
Inspect proper grounding of inner conductor.
4.
If, after 1,2,& 3 have been corrected, you still experience false counting due
to a capacitive loading effect on your test product,
change the “FD
-CO
N”
setting in the Configuration menu from “NOR”
to “RED”.
Note that your
spark tester will no longer meet the BS Standard for sensitivity but will still
meet the NEMA Standard for sensitivity.
The Fault Count
Displays “E.EXI”
1.
There is no switch or relay contact between Pins 1 & 2 (GND and HV
ENABLE). Refer to the table in “Installation” labelled “Terminal Block
Connections,” under HV ENABLE.
The Fault Count Displays “E.CV0”
1.
The clear safety cover is open.
The Fault Count Displays “E.RVF”
1.
The voltage was removed due to the spark tester being in the remove
voltage on fault mode. To
clear this press the “R” button on the front panel
.
The Fault Count Displays “E.RVC”
1.
The voltage was removed by a command through either the RS-485 serial
interface or through the fieldbus port. The voltage will need to be re-
enabled through that port.