36
Radiated
RF
IEC
61000-4-3
10V/m &
table 9
10V/m &
table 9
Portable and mobile RF commu-
nications equipment should be
used not closer to any part of the
Blood Pressure Monitor, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation dis-
tance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the com-
pliance level in each frequency
range.
b
Interference may occur in the vi-
cinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Elec-
tromagnetic propagation is affected by absorption and
reflection from structures, objects and people.