![Climax Technology ACGS-23ZW Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/climax-technology/acgs-23zw/acgs-23zw_manual_2625023003.webp)
3
Supervision Function
This function uses the Z-Wave Wake Up Command Class. The Wake Up Command Class allows the
battery-powered Glass Break Detector to notify the Control Panel/Gateway that it is awake and ready to
receive any queued commands. The wake up
time is programmed automatically according to Control
Panel’s setting when The Glass Break Detector is included. The recommended setting of the wake up
time is 60 minutes above.
Installation
The Glass Break Detector should be mounted on ceiling or wall. There should be no obstacle between
the Detector and the window protected. The Detector back cover has mounting knockouts which can
be broken for screw mounting. Alternatively, you may also mount the detector with double side
adhesive tape provided.
Mounting Location:
Mount away from sound source, such as speaker, air condition or motor.
Mount with as much distance from window or door as possible to avoid external sound interference.
Screw Mounting
1 Break through the knockouts on back cover and use them as template to mark position on wall/ceiling.
2 Drill holes at marked location and screw the back cover onto the wall/ceiling, insert wall plug if needed.
3 Replace the Detector main body onto the back cover. Make sure the cover is properly closed and the
device is fixed to the wall tightly.
Adhesive Tape Mounting
When using provided double side adhesive tape, make sure to install the detector on flat surface. Do not
install on uneven surface or location with cracking paints.
1 Clean both the detector back cover and mounting location with suitable degreaser.
2 Apply the double side adhesive tape to back cover of detector, then apply to mounting location.
<NOTE>
When mounting with double side adhesive tape, the tamper protection for mounting location
removal will be disabled. Protection from cover opening is still available.
Glass Thickness
Plated Glass: 2.4 to 6.4 mm (3/32 to 1/4")
Tempered Glass: 2.4 to 6.4 mm (3/32 to 1/4")
Wired Glass: 3.2 to 6.4 mm (1/8 to 1/4")
Laminated Glass 3.2 to 6.4 mm (1/8" to 1/4")
Testing the Detector
The Detector should be tested to ensure it is able to detect glass breakage successfully. A FlexGuard
FG-701 Glassbreak Simulator is required to test the detection function.
1 Press Function Button once to put Detector into Test mode for 5 minutes.
2 Set the FlexGuard FG-701 to “FLEX” and “TEST” mode.
3 Put FG-701 at desired testing location on the glass and point the speaker at the Detector. Close the
window covering if it is present.
4 Strike the glass with a cushioned tool to create a glass striking sound. When FG-701 detects the glass
striking sound, it will react by emitting glass break sound.
5 The Detector will only be activated if it detects both glass striking and glass break sounds. When
activated, Green LED will flash to indicate glass break has been detected.