Climax Technology ACGS-23ZW Скачать руководство пользователя страница 3

 

 

Supervision Function 

This function uses the Z-Wave Wake Up Command Class. The Wake Up Command Class allows the 
battery-powered Glass Break Detector to notify the Control Panel/Gateway that it is awake and ready to 
receive any queued commands. The wake up

 

time is programmed automatically according to Control 

Panel’s setting when The Glass Break Detector is included. The recommended setting of the wake up 
time is 60 minutes above. 

 

Installation 

The Glass Break Detector should be mounted on ceiling or wall. There should be no obstacle between 
the Detector and the window protected. The Detector back cover has mounting knockouts which can 
be  broken  for  screw  mounting.  Alternatively,  you  may  also  mount  the  detector  with  double  side 
adhesive tape provided. 

Mounting Location: 

 

Mount away from sound source, such as speaker, air condition or motor. 

 

Mount with as much distance from window or door as possible to avoid external sound interference. 

Screw Mounting 

1  Break through the knockouts on back cover and use them as template to mark position on wall/ceiling.

 

2  Drill holes at marked location and screw the back cover onto the wall/ceiling, insert wall plug if needed.

 

3  Replace the Detector main body onto the back cover. Make sure the cover is properly closed and the 

device is fixed to the wall tightly.

 

Adhesive Tape Mounting 

When using provided double side adhesive tape, make sure to install the detector on flat surface. Do not 
 install on uneven surface or location with cracking paints. 

1  Clean both the detector back cover and mounting location with suitable degreaser.

 

2  Apply the double side adhesive tape to back cover of detector, then apply to mounting location. 

<NOTE> 

 

When  mounting  with  double  side  adhesive  tape,  the  tamper  protection  for  mounting  location 
removal will be disabled. Protection from cover opening is still available. 

 

Glass Thickness 

Plated Glass: 2.4 to 6.4 mm (3/32 to 1/4") 
Tempered Glass: 2.4 to 6.4 mm (3/32 to 1/4") 
Wired Glass: 3.2 to 6.4 mm (1/8 to 1/4") 
Laminated Glass 3.2 to 6.4 mm (1/8" to 1/4") 

 

Testing the Detector 

The Detector should be tested to ensure it is able to detect glass breakage successfully. A FlexGuard 
FG-701 Glassbreak Simulator is required to test the detection function. 

1  Press Function Button once to put Detector into Test mode for 5 minutes.

 

2  Set the FlexGuard FG-701 to “FLEX” and “TEST” mode.

 

3  Put FG-701 at desired testing location on the glass and point the speaker at the Detector. Close the 

window covering if it is present.

 

4  Strike the glass with a cushioned tool to create a glass striking sound. When FG-701 detects the glass 

striking sound, it will react by emitting glass break sound.

 

5  The Detector will only be activated if it detects both glass striking and glass break sounds. When 

activated, Green LED will flash to indicate glass break has been detected.

 

 

 

 

Отзывы: