25
3. Apply Control Solution
Shake the control solution vial thor-
oughly before use.
Squeeze out a drop and wipe it off, then
squeeze out another drop and place it
on the tip of the vial cap.
Hold the meter to move the absorbent hole of the test strip to touch
the drop. Once the confirmation window fills completely, the meter
will begin counting down.
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
4. Read and Compare the Result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and
it should fall within this range. If not,
please read the instructions again and
repeat the control solution test.
Normal
109~165