Clements SUC89851 Скачать руководство пользователя страница 4

Installation and use 

 

 WARNING 

FOR THE CORRECT USE OF THE COLLECTION CONTAINER, READ THE INSTRUCTIONS IN THE USER 

MANUAL SUPPLIED WITH THE DEVICE. 

THE LIQUIDS COLLECTED MUST ONLY BE DRAINED OUT IN AREAS ASSIGNED TO THE DISPOSAL OF 

HOSPITAL WASTE AND/OR STRICTLY FOLLOWING THE INSTRUCTIONS OF THE AUTHORITIES FOR THE 

TREATMENT OF THESE PRODUCTS.

 

Cleaning and disposal  

 

 

Clean the device every day or in accordance with the hospital routine 

 

Clean or replace the sintered bronze filter (6) if necessary. 

 

Carefully clean all the external surfaces of the device using a soft damp cotton cloth rinsed in neutral detergent 

diluted 10% in water. 

  

CAUTION

 

DO NOT USE SOLVENTS OR ABRASIVE PRODUCTS FOR CLEANING: THESE CAN SERIOUSLY DAMAGE 

THE SURFACES OF THE EQUIPMENT AND THE PLASTIC PARTS. 

 

DO NOT IMMERSE THE UNIT IN DISINFECTANT; 

 

DO NOT PLACE THE UNIT IN AN AUTOCLAVE; 

 

DO NOT USE INFLAMMABLE PRODUCTS. 

IF THE EQUIPMENT IS ACCIDENTALLY CONTAMINATED BY SUCTIONED LIQUIDS, IT MUST BE ISOLATED 

AND THE NECESSARY CLEANING AND DISINFECTION OPERATIONS MUST BE CARRIED OUT BY 

QUALIFIED PERSONNEL. 

IF NECESSARY, STERILISE THE VENTURI  NOZZLE  WITH A VALIDATED  ETHYLENE OXIDE 

STERILIZATION CYCLE (E.G. 14% EO 12 H. / 50 KPA / 37 °C). 

  

CAUTION FOR DISPOSAL

 

IF THE DEVICE IS DAMAGED AND CANNOT BE REPAIRED, DISPOSE AND DISCARD IT ACCORDING TO 

THE INTERNAL PROCEDURES OF THE HOSPITAL. DO NOT RECYCLE OR REUSE ANY COMPONENT.

 

 

Technical features 

Sales description ..................................................  

 

Dimensions (mm) ..................................................  

 

 

 

Weight (g) .............................................................  

 

Scale of the vacuum gauge ...................................  

 

 

Maximum suction flow ...........................................  

 

Maximum supply depression .................................  

 

Vacuum supply connection ...................................  

 

Vacuum connector ................................................  

 

Storage conditions ................................................  

 

Operating conditions .............................................  

Direct suction nozzle 

 

Height: 195  

Width: 68  

Depth: 48 

 

400 

 

From 0 to -100 kPa graduated at 5 kPa 

From 0 to -760 mm Hg graduated at 20 mm Hg 

 

60 L/min ± 5 L/min at -98 kPa 

 

-98 kPa 

 

Thread M8x1 F 

 

Hose connector Ø 10÷12mm 

 

-40 °C ± 2 °C / +60 °C ± 5 °C and 40% to 70% 

RH

 

 

+5 °C ± 2 °C / +35 °C ± 5 °C 