background image

 

CMS-2 CARBON MONOXIDE MONITOR 

Page 2 

 

 

 

 

 

© 2018 CLEMCO INDUSTRIES CORP.  

  www.clemcoindustries.com  

  Manual No. 22925, Rev. I, 04/18 

SERVICE MAINTENANCE  ......................................... 6.0 

Sensor Replacement  ................................................... 6.1 
Fuse Replacement  ...................................................... 6.2 

 

TROUBLESHOOTING  ................................................ 7.0 

Instrument Will Not Calibrate  ...................................... 7.1

 

Intermittent Alarm  ........................................................ 7.2 
No Annunciator Lights or Alarms  ................................ 7.3 
Either Alarm Lights or Alarm Horn Fails  ...................... 7.4 

 

ACCESSORIES AND REPLACEMENT PARTS  ....... 8.0 
 
1.4 Components 

 

 

1.4.1 

The primary components of the monitor are 

shown in Figure 1. The instrumentation is enclosed in a 
corrosion resistant polypropylene carrying case. The 
monitor system includes the instrument case, 10 feet of 
1/8" ID x 1/4" OD tubing with fittings, a cylinder of 10 ppm 
(parts per million) test gas, calibration connector that 
includes the connector valve, tubing, and humidifier. 
 

1.5 Operating 

Principles 

 

1.5.1

 

The CMS monitor samples respiratory air from a 

breathing-air source. This enables one monitor to detect 
CO and other toxic gases in the entire breathing air 
system, not just for one respirator. The air to the 
respirator does not actually pass through the monitor.  

 

 

WARNING 

When the CMS is correctly installed and 
maintained, it monitors the level of CO and 
other oxidizable toxic gases in the air supply. 
The instrument is a monitoring device; it DOES 
NOT remove or convert CO or any other toxic 
gases.  

 
 

 
 
 

 

 

 

Figure 1 

 

CLEMCO

 

CMS-

2

10 ppm CO 

or 

Impurity-Free

 

Test Gas

 

Fuse 

Audible Alarm 

Case Cover 

Test Gas Port 

Flow Meter 

Cover Screw 

Case Cover Latch 

Tubing

1/4

"

 OD Tube Fitting (2)

This end connects to air source.

Calibration Connector 
with Flow-Control Valve 

Humidifier   

Run/Calibration Switch 

Calibration Adjustment 

Calibration Indicator 

Exterior Alarm Light 

120-Volt Power Cord 

Calibration Connector Tube End

Plugged Port 
for Optional Remote Alarm 

Содержание CMS-2

Страница 1: ...e WARNINGS are included for the health and safety of the operator and those in the immediate vicinity Electronic files include a Preface containing the same important information as in the orange cover WARNING 2018 CLEMCO INDUSTRIES CORP One Cable Car Dr Washington MO 63090 Phone 636 239 4300 Fax 800 726 7559 Email info clemcoindustries com www clemcoindustries com ...

Страница 2: ...tion that if not avoided could result in minor or moderate injury WARNING Warning indicates a hazardous situation that if not avoided could result in death or serious injury DANGER Danger indicates a hazardous situation that if not avoided will result in death or serious injury 1 3 Table of Contents SUBJECT SECTION LOCATION INTRODUCTION 1 0 Scope of Manual 1 1 Safety Alerts 1 2 Table of Contents 1...

Страница 3: ...ier 1 5 Operating Principles 1 5 1 The CMS monitor samples respiratory air from a breathing air source This enables one monitor to detect CO and other toxic gases in the entire breathing air system not just for one respirator The air to the respirator does not actually pass through the monitor WARNING When the CMS is correctly installed and maintained it monitors the level of CO and other oxidizab...

Страница 4: ...ion Do not use the compressed air for breathing unless the exterior light is illuminated green Yellow Light Intermittent Alarm Horn approximately every 10 seconds Sample Air Failure This alerts the user that pressure in the test chamber is either too low or too high Low pressure is usually due to loss of pressure in the sample air air line High pressure is an indication that the internal regulator...

Страница 5: ...d to seal pipe fittings The monitor will respond to the gases given off by these compounds Use Teflon tape for pipe thread sealer Do not use any silicone based lubricants up stream of this instrument Silicone exposure will prematurely damage the sensor 2 3 1 To ensure accurate and rapid air monitoring the monitor must be placed as close to the compressed air source as practical Ten feet of 1 8 ID ...

Страница 6: ... AC to DC field conversion Serial No 4446 and greater To convert monitors with serial numbers lower than 4446 request Manual No 22925 Revision E or earlier The AC power cord provided with the monitor may be used using black as the positive A user supplied DC 2 wire cord may be used using red as positive If a new 12 volt cord is used splice the positive through the fuse and use the spade on the out...

Страница 7: ...o operate relays for external alarms and shut down devices 2 6 Changing Mode Select Dip Switch to Delay Alarm Figure 5 NOTE In areas where intermittent radio frequency transmissions cause false alarms the monitor can be set to have a 10 second alarm delay to avoid false alarm conditions The dip switch is mounted on the circuit board as shown in Figure 5 WARNING Do not change the switch positions u...

Страница 8: ...Section 3 3 3 2 4 If the instrument goes into an alarm exterior light remains red with continuous audible alarm after the warm up reinitiate warm up by terminating and restarting power 3 2 5 Continued alarm condition after several warm up periods indicates that the sample air exceeds the permissible contamination level or the instrument requires calibration Refer to Section 4 0 WARNING DO NOT use ...

Страница 9: ...river when calibration is required 4 3 Stabilization Period 4 3 1 The monitor must be in operation for at least 30 minutes before calibration tests DO NOT test the instrument until it has gone through the warm up period and has been in operation with sample air flowing through it see Section 3 0 for at least 30 minutes 4 4 Prepare Calibration Connector NOTE Unlike ambient air calibration test gas ...

Страница 10: ...un Calibration toggle toward CALIBRATION The external alarm light will immediately turn yellow and within a few seconds the intermittent audible alarm will sound the calibration indicator remains green 4 6 3 With the flow meter vertical refer to Figure 9 and slowly open the flow control knob to introduce test gas Test gas is entering the unit when the flow meter ball rises If the ball does not ris...

Страница 11: ...r light turns red 4 7 5 Once the indicator is red turn the potentiometer in the opposite direction until the light turns green Then slowly turn the potentiometer from green to red several times to find the trip point The monitor is calibrated at the spot where the light turns red Return the instrument to operating mode per Section 4 8 4 8 Return Monitor to Operating Mode 4 8 1 Turn the flow contro...

Страница 12: ...death 5 0 GENERAL MAINTENANCE 5 1 Care and Handling 5 1 1 This monitor is designed for portable field use and is not adversely affected by normal handling that is required of any test and measurement instrument 5 1 2 Transport the instrument in the passenger compartment of the service vehicle NOTICE Do not subject this instrument to extreme heat or cold Placing the instrument on the dashboard of t...

Страница 13: ...b as the new sensor must be installed in the same orientation While supporting the circuit board carefully pull straight up to remove the sensor 6 1 9 If the sensor comes off with the housing note the tab location and remove the sensor by pulling on a couple of the sensor pins 6 1 10 Discard the old sensor to avoid attempts to reuse 6 1 11 Handle the new sensor taking care not to distort the pins ...

Страница 14: ...sengage as soon as the pressure is corrected After the pressure is correctly set close the faceplate 7 2 4 If the flow meter ball does not raise or stays very low in the meter 7 2 4 1 Check internal instrumentation lines for breaks kinks or disconnection If a line has come loose reconnecting it will re establish function and the monitor can be correctly used NOTE This type of failure could be the ...

Страница 15: ... 120 volt AC 22892 12 volt DC 25072 120 volt AC monitors can be field converted to 12 volt DC Refer to Section 2 4 1 Calibration connector assembly includes tube humidifier and flow control valve with connector 22893 2 Test gas 10 ppm 22865 Impurity free 11132 3 Tubing clear for Item 1 and internal per ft specify length required 29261 4 Regulator filter M5 22920 5 Filter element replacement for It...

Отзывы: