There are 2 UVA lamps and 2 UVC lamps installed in the neutralization chamber. After starting the
fan, the first pair of lamps /UVA+UVC/. After 60 seconds, the second pair of lamps starts up. Lamp
operation time is saved in the device memory. The standard life of each lamp is 9,000 hours.
ATTENTION
•
total working time of the lamps is to be reset in service menu after lamp
replacement;
•
lamps must be replaced all 4 at the same time.
5.3
Fan module
Figure 16
The fan module (Figure 16) is located in the upper part of appliance. All components are individually
replaceable.
The fan module consist of:
1.
Top cover open limit switch; this signals to the system whether the cover is open.
2.
HEPA H14 filter locks, this locks the filter when installed.
3.
Main HEPA H14 filter, this cleans the air.
4.
Fan, this is responsible for the air flow through the device.