Cleaver Scientific MSSCREEN-TRIO Скачать руководство пользователя страница 14

15/11/2018 

 

Page 12 

Troubleshooting 

Problem 

Cause 

Solution 

Bands sharp but not enough 
bands seen 

Gel agarose percentage too 
high 

Incomplete digestion 

Decrease agarose percentage. 

Review enzyme activity, digest 
further. 

Band smearing and streaking 

Agarose has improper 
endosmosis 

Salt concentration in sample too 
high 

Excessive power and heating 

Sample spilled out of well 

Incomplete digestion, nuclease 
contamination, bad enzyme 

Sample wells cast through the 
gel. Sample leaks along bottom 
of running surface 

Consult Cleaver Scientific about 
agarose. 

Reduce salt concentration to 

≤0.1M. 

Reduce voltage. See 
electrophoresis instructions. 

Apply sample carefully. Increase 
gel thickness for large sample 
volumes. Adjust comb height. 

Heat sample. Review enzyme 
activity. Digest sample further. 

Comb should be placed to 1 to 2 
mm above the base of the 
running surface. 

Curved line or distortion of bands 

Bubbles in sample wells 

Remove bubbles prior to 
electrophoresis. 

Curved bands, smiles 

Sample overload 

Reduce load. 

Differential relative mobilities 

Sample spilled out of wells 

Unit not levelled 

Samples should have proper 
density. Apply carefully. 

Level unit. Use a steady work 
bench. 

Gels crack 

Too high voltage gradient, 
especially with low melting 
temperature agarose or low gel 
strength gels 

Reduce voltage. Run gel at lower 
temperature. 

High MW bands sharp; low MW 
bands smeared 

Gel agarose percentage too low 

Increase agarose percentage. 
Switch to polyacrylamide. 

Ragged bands 

Sample density incorrect 

Sample well deformed 

Excessive power or heating 

See sample application 
instructions. 

Carefully remove comb, 
especially from soft gels. Make 
sure gel has solidified. 

Cooling soft gels aids in comb 
removal. 

Reduce voltage. See 
electrophoresis instructions. 

Slanted lanes (bands) 

Gel not fully solidified 

Comb warped or at an angle 

Gel to solidify for at least 30-
45min. 

Check alignment of comb. 

Содержание MSSCREEN-TRIO

Страница 1: ...multiSUB Screen Horizontal System...

Страница 2: ......

Страница 3: ...RIO MSSCREEN16 NC MSSCREEN24 NC MSSCREEN32 NC MSSCREEN TRIO NC Record the following for your records Model _____________________ Catalogue No _____________________ Date of Delivery ___________________...

Страница 4: ...ons 6 Setting up the Horizontal Gel Tanks 6 Fitting Electrodes 6 Fitting Loading Guides 6 Gel Preparation 7 Gel Pouring 8 Flexicaster 8 Tape 9 Running the Gel 9 Gel Staining and Viewing 10 Solutions 1...

Страница 5: ...he complete manual thoroughly The unit must never be used without the safety lid correctly in position The unit should not be used if there is any sign of damage to the external tank or lid These unit...

Страница 6: ...x 16cm W x L 6 x MS26 28MC 1 1mm 28 sample MS26 LG Strips MS26 WP Platform CSL CAB None MSSCREEN24 NC MS26 UV24 26 x 24cm W x L 6 x MS26 28MC 1 1mm 28 sample MS26 LG Strips MS26 WP Platform CSL CAB No...

Страница 7: ...ions Power supply cables 1500V rated with retractable 4mm connectors Safety lid and viewing pane High impact acrylic construction Height adjustable combs Acrylic UV transparent gel tray UV transparent...

Страница 8: ...Fitting Electrodes Note the position of the lid on the unit This shows the correct polarity and the correct orientation of the cables black is negative and red positive Remove the lid from the unit N...

Страница 9: ...ose powder to a conical flask 2 Add the appropriate amount of 1x TAE or TBE solution from the table above To prevent evaporation during the dissolving steps below the conical flask should be covered w...

Страница 10: ...end of the Flexicaster 3 Position the movable end of the Flexicaster so that the silicone mat is pushed against the other end of the tray 4 Turn the cam so that the silicone mat tightly seals against...

Страница 11: ...ip 4 Allow the agarose to set ensuring that the gel remains undisturbed 5 Carefully remove the gel casting gates and comb and transfer the gel including tray to the main tank Running the Gel 1 Mix the...

Страница 12: ...sed 3 Rinse the gel twice for a couple of seconds with distilled water 4 Transfer the gel to a UV Transilluminator The samples will often appear as brighter clearer bands when photographed or viewed u...

Страница 13: ...lycerol 0 25 bromophenol blue and 0 25 xylene cyanole FF in 1x TAE buffer Only 1 10 ml of the 10x loading dye should be prepared Ethidium Bromide Solution Add 10 mg of Ethidium Bromide to 1 ml distill...

Страница 14: ...running surface Curved line or distortion of bands Bubbles in sample wells Remove bubbles prior to electrophoresis Curved bands smiles Sample overload Reduce load Differential relative mobilities Sam...

Страница 15: ...gents include dishwashing liquid Hexane and Aliphatic hydrocarbons The units should not be left to in detergents for more than 30 minutes The units should never come into contact with the following cl...

Страница 16: ...C 0 75mm thick MS26 56MC 0 75 Comb 56 sample MC 0 75mm thick MS26 28MC 1 Comb 28 sample MC 1mm thick MS26 56MC 1 Comb 56 sample MC 1mm thick MS26 28MC 1 5 Comb 28 sample MC 1 5mm thick MS26 56MC 1 5 C...

Страница 17: ...5g tablets CSL RUNSAFE CSL RUNSAFE Package 1 ml vial CSL TBEP Powdered Tris Borate EDTA Running Buffer 10 sachets 1litre pk TBE10X1L Cleaver Buffer Tris Borate EDTA Running Buffer 10 x 1L TBE10X5 Cle...

Страница 18: ...n performed by anyone other than CSL or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units which have accessories or repaired parts not s...

Страница 19: ......

Страница 20: ......

Отзывы: