U-PAD2 A/D Converter
5 Troubleshooting
5.5 Data Acquisition - Simulated
▌ The title "Simulated" is displayed in the
window.
Fig 40: Data Acquisition Simulated
Description:
The corresponding Instrument only displays the simulated curve (from the file
CHANNX.DTA file) in the
window, or in other words has allocated
a
DEMO detector
on itself.
Solution:
Open the
dialog from the
window using the
System -
Configuration...
command and check the tab of the corresponding Instrument -
Instrument X
. If it only has the detector signals from the
DataApex DEMO
detector, it is necessary to reconfigure it. You can find more information how to add
a detector in the chapter
..
Make sure that you have installed a full version of
Clarity
. If this is not the case,
inscription
DEMO
is in the header of the main window. If so, install the full version
and enter the user code you have been provided. Also make sure that the detector
is properly configured.