Filter Life
GUS® has a patented filter that, in normal everyday use, will have a six-month life. Call CIVCO Medical Solutions at
1-800-445-6741 to purchase additional filters.
Filter Replacement
WARNING: Put cap on the tube containing your disinfectant. Turn the main power OFF and disconnect
the power cord from the electrical outlet.
1. Remove the foam pad from the top compartment.
2. Remove the top plate located in the top compartment by turning the captive screws counterclockwise.
3. Remove the old filter by inserting fingers into the finger holes and pulling up. Filter may be snug but should
slide up with steady pressure.
4. Using a mild detergent on a soft cloth, wipe down the whole front of the machine. Also, wipe inside the
machine behind the air intake panel. Dry all surfaces.
5. Replace with a new filter, making sure the airflow arrows are pointing to the rear of the machine.
6. Put the used filter into a plastic bag and dispose of it with other non-hazardous waste.
This does not need to
be red tagged.
7. Replace the top plate and tighten the captive screws by turning clockwise. Replace the foam pad and turn on
the machine before replacing disinfectant.
Cleaning the System
WARNING: Turn the main power switch OFF and disconnect the power cord from the electrical outlet.
DAILY CLEANING:
1. Remove the rinse water container, empty and wash both the container and cap. Some facilities do this after
each use.
2. Using a mild detergent solution on a soft cloth, wipe down the area around the containers.
3. Wipe off any drips around the disinfectant container and cap and below the shelf.
4. Do not spray cleaning liquids on the intake area of the machine as this will damage the filter.
5. After drying the areas, replace the rinse water and disinfectant containers.
Refer to
“Filter Replacement”
section for additional cleaning steps when changing the filter.
M0046_H
9
G10VPL
GUS Disinfection Soak Station™