Filter Life

GUS® has a patented filter that, in normal everyday use, will have a six-month life. Call CIVCO Medical Solutions at
1-800-445-6741 to purchase additional filters.

Filter Replacement

WARNING: Put cap on the tube containing your disinfectant. Turn the main power OFF and disconnect
the power cord from the electrical outlet.

1. Remove the foam pad from the top compartment.

2. Remove the top plate located in the top compartment by turning the captive screws counterclockwise.

3. Remove the old filter by inserting fingers into the finger holes and pulling up. Filter may be snug but should

slide up with steady pressure.

4. Using a mild detergent on a soft cloth, wipe down the whole front of the machine. Also, wipe inside the

machine behind the air intake panel. Dry all surfaces.

5. Replace with a new filter, making sure the airflow arrows are pointing to the rear of the machine.

6. Put the used filter into a plastic bag and dispose of it with other non-hazardous waste.

This does not need to

be red tagged.

7. Replace the top plate and tighten the captive screws by turning clockwise. Replace the foam pad and turn on

the machine before replacing disinfectant.

Cleaning the System

WARNING: Turn the main power switch OFF and disconnect the power cord from the electrical outlet.

DAILY CLEANING:

1. Remove the rinse water container, empty and wash both the container and cap. Some facilities do this after

each use.

2. Using a mild detergent solution on a soft cloth, wipe down the area around the containers.

3. Wipe off any drips around the disinfectant container and cap and below the shelf.

4. Do not spray cleaning liquids on the intake area of the machine as this will damage the filter.

5. After drying the areas, replace the rinse water and disinfectant containers.

Refer to

“Filter Replacement”

section for additional cleaning steps when changing the filter.

M0046_H

9

G10VPL

GUS Disinfection Soak Station™

Содержание GUS G10VPL

Страница 1: ...741 WWW CIVCO COM Copyright 2018 All Rights Reserved GUS is a registered trademark of CIVCO Medical Solutions All other trademarks are property of their respective owners Printed in the USA G10VPL Wal...

Страница 2: ...Mounting Instructions 4 Using High Level Disinfectants with GUS 6 Step 1 Probe Soaking Instructions 6 Step 2 Filling the Containers 7 Step 3 Ready for Use 7 Repair and Modification 8 Filter Life 9 Fi...

Страница 3: ...ed Safety Precautions Disconnect power before replacing filter or when servicing the unit Refer to the MSDS of the high level disinfectant that you are using for additional safety precautions Environm...

Страница 4: ...inst unsafe practices or potential equipment damage NOTE Indicates additional helpful information Quick Troubleshoot If the GUS system is not running check the following 1 Ensure that the unit is plug...

Страница 5: ...ns This MUST be placed above the workstation l One Bottom support bracket l One filter installed in unit Wall Mounting Instructions 1 You will need the following items before attempting to install thi...

Страница 6: ...p 5 9 Place the unit back over the two top screws Screw the two remaining Philip screws into the bottom bracket and tighten Tighten the top screws CAUTION Do not over tighten the screws 10 For masonry...

Страница 7: ...ble as follows If your probe is NOT water resistant Suspend the probe into an empty container so that the top of the handle is level with the top of the container Depress blue tab and place probe cabl...

Страница 8: ...as there are chemicals inside the containers 5 Fill the remaining container with clean rinse water for the preliminary rinse You MUST have one container for rinse water This ensures that when the pro...

Страница 9: ...complete push back on the blue tab and lift the probe up shake it gently to remove large drops of disinfectant Then place it directly into the rinse container while keeping your thumb on the red mark...

Страница 10: ...ting to the rear of the machine 6 Put the used filter into a plastic bag and dispose of it with other non hazardous waste This does not need to be red tagged 7 Replace the top plate and tighten the ca...

Страница 11: ...spills CIVCO Medical Solutions manufactures Glute Out a glycine based neutralizer part 610 1099 which comes in a convenient single use bottle Glute Out is specifically for use with glutaraldehyde and...

Страница 12: ...BILITY AND OF FITNESS FOR A PARTICULAR PURPOSE ARISING FROM A COURSE OF DEALING OR USAGE ARE SPECIFICALLY EXCLUDED This warranty does not cover conditions and damages resulting from any of the followi...

Отзывы: