ASTRA® VR Endovaginal/Endorectal Probe Reprocessor
Installation Manual
M0065_C
4
ASTRA® VR
5.
INSTALLATION
A.
Typical Installation
The ASTRA VR is a floor standing unit which requires a potable water supply, waste drain, and
GFCI electrical connection all within 8ft (2.4 m) of the installation location. The ASTRA VR
device is required to be mounted to the wall using a supplied wall bracket. Failure to follow
these installation instructions may affect the operation and effectiveness of this device.
Figure 1 below shows a typical installation.
Figure 1
Typical ASTRA VR Installation
The on/off switch, power cord, vent, water supply, and drain are located on the right side
(facing) of the device. These connections may be routed behind the unit if your plumbing or
electrical outlet is located to the left. Adequate space is required to ensure proper safe
access to controls and proper ventilation of the device. See Section 3 for detailed
installation and clearance requirements.