CLEANING INSTRUCTIONS
The CTD505 does not require a probe cover during use. Clean and disinfect the
CTD505 before and after each use. If you wish to share the CTD505 with others,
follow the instructions below.
◆
Preparation
1. Cotton bud: Available in supermarkets.
2. 70 % concentration alcohol swab: Available in
supermarkets.
Note: Dispose of used cotton buds and alcohol swabs in
accordance with local regulations.
◆
Cleaning : Probe area
Use a clean cotton bud to wipe off any visible dirt or ear wax. Gently wipe the area
clean.
Note: The most delicate and sensitive part of the CTD505 is the sensor inside the
probe tip. Poking it with force may cause damage to the CTD505. When cleaning,
take extra care.
Hold the CTD505 facing down to prevent water getting into the CTD505. Use an
alcohol swab to clean the probe area in a circular motion for 15 seconds (around 20
times).
DISINFECTION (Probe & Exterior Area)
1. Hold the CTD505 facing down to prevent water getting
into the CTD505. Use an alcohol swab to clean the probe
area in a circular motion for 15 seconds (around 20
times).
2. Then, use the same alcohol swab to wipe the exterior
area back and forth for another 15 seconds (around 20
times).
3. Wait at least 30 seconds until the CTD505 is completely
dry.
Continue to set units on the next page, and start measuring.
Repeat the steps above when measurement is done.
NOTE:
Do not use detergent or any strong chemicals on the CTD505.
FEVER ALERT
The CTD505 has a built-in “Fever Alert” feature.
If you have a fever between 37.5 °C – 43.0 °C, the “Fever Alert” function will inform
you by a fl ashing result, red LED backlight, and a series of beeps.
NOTE:
Do not use the CTD505 as a replacement for medical advice. Consult your doctor if
you have any concerns.