- Eng 20 -
English
Icon displayed
Condition/Cause
Displayed during
measurement.
Lit
Measurement is under way.
Flashing
Pulse detected.
Displayed during
measurement.
“PUL Err” is displayed when the pulse
value is below 40 or above 181.
If any of the following marks is displayed, take the specifi ed action, then
repeat the operation.
Icon displayed
Condition/Cause
Corrective action
• Insuf
fi
cient pressure.
• Appears
when
the
blood-pressure could
not be measured
accurately.
Apply cuff properly and
measure again.
The pressure is 281
mmHg or higher.
If the pressure does not
automatically decline
during measurement,
immediately press the
“START/STOP” switch
and turn the unit off. To
take the measurement
again, refer to P.16
and correctly re-apply
pressure.
Appears when the
batteries are running low.
Replace the batteries.
Detected problem in the
device.
Contact the vendor.
KEY TO DISPLAY ICONS