After setting up the Doppler String Phantom, turn on the power switch (located on the back of the controller). 
The display reads 

UP=H2O  DOWN=H2O/GLYCOL

1.  If the tank is filled with water, press 

UP

. If the tank is filled with velocity-corrected fluid, press 

DOWN

 

(See About Velocity-Corrected Fluid on page 4 for more information.)

 

The display reads 

SELECT A FUNCTION

.

2. Press 

TEST

 or 

PHYSIOLOGICAL

 to select a waveform group.

 

The display shows a waveform.

3. Press 

UP

 or 

DOWN

 to select a waveform. (See Appendix A for graphs of waveforms.)

4. Press 

EXECUTE

 to start the motor. 

5.  To select a different waveform, you may press 

UP

 or 

DOWN

 at any time while the motor is running. 

6.  To reverse the direction of simulated flow, press 

FLOW REV

 at any time. 

 

An arrow in the bottom right corner of the display indicates the flow direction. 

7.  To change functions (for example, to switch from physiological waveforms to constant speeds), press    
 

ESCAPE

.  The display reads 

SELECT A FUNCTION

.

RS-232 INSTRUCTIONS

Introduction

The remote control software allows all of the same functions that are available from the front panel.  It is primarily 
a serial communications feature for use with automatic testing equipment, but may be used by anyone wanting 
to control the phantom for standardized testing.  It may be controlled by a laptop, desktop, or a process control 
computer.

To establish the link between the Doppler String Phantom controller and your computer, any common RS-232 
communications program such as Pro-Comm Plus, Teraterm Plus, Microsoft Hyperterminal, etc., may be used.  
The communications program must be set to 9600, N, 8, 1.  Use hardware flow control.  If your computer does 
not have an RS-232 port, you can use an adapter available locally that will convert a standard USB port to RS-
232.

Getting Started

To begin using the remote control feature:

1.  Ensure the Doppler String Phantom controller is powered off (rear panel switch).

2.  Plug the power cable into the 

AC Receptacle

 and into an AC power source.  The label on the AC  

 

Receptacle  indicates 120 volts. If using with 220 volts, use step down controller (Model 901-C, sold sepa- 

 rately).

3.  Plug the custom RS-232 cord (supplied with the Doppler String Phantom) into the 

RS-232

 receptacle  

 

on the back of the controller.

4.  Plug the motor cable into the 

Drive Motor

 receptacle on the back of the controller.

5.  Start the communications program.

6.  Turn on the power on the controller (rear panel switch).

Содержание Doppler String Phantom 043

Страница 1: ...900 Asbury Ave Norfolk Virginia 23513 USA Tel 757 855 2765 WWW CIRSINC COM USER GUIDE Doppler String Phantom Model 043 ...

Страница 2: ......

Страница 3: ... 7 Getting Started 7 Accessing the Help Menu 8 Selecting Velocities and Waveforms 9 Setting Up for Quality Control Testing 9 Locking the Settings 10 Description of Commands Available Via RS 232 10 MAINTENANCE INSTRUCTIONS 11 Making String Loops 11 Ordering String Loops 11 Cleaning the Doppler String phantom 11 Calibrating the Doppler String Phantom 11 Replacing a fuse 12 ACCESSORIES AND PARTS 12 C...

Страница 4: ...ary tap water It is simple to use and almost no maintenance is required An RS 232 interface is also provided for automated use if desired Ultrasound technologists biomedical technicians researchers and manufacturers all rely on the Doppler String Phantom for hassle free reliable testing Starting a Quality Assurance Program Each ultrasound lab will have different Quality Assurance QA needs dependin...

Страница 5: ...should find Electronic Controller Acrylic tank with motor and pulleys attached Scanhead Clamp Set of 5 String Loops not pictured Rubber Diffuser Mat not pictured Power Cord not pictured Carrying Case not pictured Figure 1 Doppler String Phantom Electronic Controller Scanhead Clamp Motor Acrylic Tank ...

Страница 6: ...ow the motor and around the pair of Adjustable Pulleys at the other end of the tank See Figure 2 4 Adjusting the string tension by moving the Adjustable Pulleys until the string is fairly tight but not tight enough to break The actual tension will not affect the data Figure 2 String Installation About Reverberations Water conducts ultrasound more efficiently than the human body so it may be necess...

Страница 7: ...rarily it should be stored in a dry location to prevent corrosion Figure 3 Scanhead Clamp Clamp Adjusting Knob About Velocity Corrected Fluid The speed of sound is 1480 meters per second in water and 1540 meters per second in the human body Doppler scanners are calibrated for 1540 meters per second so some ultrasound profession als prefer to use a water gly col mix that also has a speed of sound o...

Страница 8: ...nel of the phantom This is a rarely used feature that can aid ultrasound researchers as they test new ultrasound machines Since most users of this phantom do not deal with that aspect it may be simply ignored When used this output will generate a signal pulse at the very beginning of each test or physiological waveform That pulse is 10 milliseconds long and rises from zero volts up to five volts a...

Страница 9: ...everse the direction of simulated flow press FLOW REV at any time An arrow in the bottom right corner of the display indicates the flow direction 5 To change functions for example to switch from constant speeds to test waveforms press ESCAPE The display reads SELECT A FUNCTION Using Preprogrammed Waveforms The Doppler String Phantom has several preprogrammed waveforms to help you test your ultraso...

Страница 10: ...marily a serial communications feature for use with automatic testing equipment but may be used by anyone wanting to control the phantom for standardized testing It may be controlled by a laptop desktop or a process control computer To establish the link between the Doppler String Phantom controller and your computer any common RS 232 communications program such as Pro Comm Plus Teraterm Plus Micr...

Страница 11: ... When the text appears the system is ready to test Notes When entering commands either upper or lower case may be used Upper case is used in this manual for easier reading Between each command wait until the controller responds that it s ready for the next command Accessing the Help Menu The Help Menu shown below can be displayed at any time by entering Ctrl H The Help Menu is also displayed autom...

Страница 12: ... a human Doppler waveform such as the Common Carotid waveform the following steps must be taken 1 Turn the controller on or enter Ctrl Z to restart the program 2 Issue U or D to select the type of fluid in the tank 3 Issue P to select Physiological waves The screen displays Common Carotid 4 To select a different waveform issue U to move up the list Continue issuing U or D until the desired wavefor...

Страница 13: ...example S52 Enter Decimal numbers are not allowed T TEST WAVES Used to access test waveforms P PHYSIOLOGICAL WAVES Used to access physiological waveforms U UP KEY Used to select water type in tank move up in the menus and to increase constant speeds D DOWN KEY Used to select water type in tank move down in the menus and to decrease constant speeds E EXECUTE KEY Used to select options currently dis...

Страница 14: ...ing ends allow it to dry slightly and then roll the knot between your fingers to form a smooth ball Ordering String Loops You can order a pack of 20 preformed string loops made from 3 0 surgical silk The Part number is 043SL Please call for current prices Cleaning the Doppler String Phantom The Doppler String Phantom is designed to be virtually maintenance free You can clean the controller and tan...

Страница 15: ...e used with the Doppler String Phantom when using outside of the USA The Part number is 043C Please call for current pricing How to replace a fuse The Doppler String Phantom contains a fuse that may need to be replaced periodically If the unit does not power up first check the cord and power source If all of the connections are properly in place then the fuse may need to be replaced Unplug the uni...

Страница 16: ...tor prior to returning the product Visit https www cirsinc com distributors to find your local distributor If you purchased your product direct through CIRS call Customer Service at 800 617 1177 email rma cirsinc com or fax an RMA request form to 757 857 0523 CIRS staff will attempt to remedy the issue via phone or email as soon as possible If unable to correct the problem a return material author...

Страница 17: ...14 APPENDIX A GRAPHS OF WAVEFORMS TEST WAVEFORMS Sine Wave 100 cm sec ...

Страница 18: ...15 APPENDIX A GRAPHS OF WAVEFORMS Sine Wave 150 cm sec ...

Страница 19: ...16 Sine Wave 200 cm sec APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 20: ...17 Triangle Wave 100 cm sec Triangle Wave 150 cm sec APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 21: ...18 Triangle Wave 200 cm sec Stepped Ramp APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 22: ...19 Adult Aortic APPENDIX A GRAPHS OF WAVEFORMS PHYSIOLOGICAL WAVEFORMS Adult Common Carotid ...

Страница 23: ...20 Adult Femoral APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 24: ...21 Pediatric Umbilical Stenotic Common Carotid APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 25: ...22 Pediatric Renal APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 26: ...23 Pediatric Ductus Arteriosus APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 27: ...24 Pediatric Middle Cerebral APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 28: ...25 Pediatric Descending Thoracic APPENDIX A GRAPHS OF WAVEFORMS ...

Страница 29: ...26 APPENDIX B SAMPLE QA FORM Sample ...

Страница 30: ...27 APPENDIX B SAMPLE QA FORM ...

Страница 31: ......

Страница 32: ...t notice Publication 043 UG 072020 Computerized Imaging Reference Systems Inc has been certified by UL DQS Inc to ISO 13485 2016 Certificate Registration No 10000905 MP2016 COMPUTERIZED IMAGING REFERENCE SYSTEMS INC 900 Asbury Ave Norfolk Virginia 23513 USA Toll Free 800 617 1177 Tel 757 855 2765 Fax 757 857 0523 Email admin cirsinc com www cirsinc com Technical Assistance 1 800 617 1177 ...

Отзывы: